    Design and In Silico Evaluation of a Closed-Loop Hemorrhage Resuscitation Algorithm With Blood Pressure as Controlled Variable

    Source: Journal of Dynamic Systems, Measurement, and Control:;2021:;volume( 144 ):;issue: 002::page 21001-1
    Author:
    Alsalti, Mohammad
    ,
    Tivay, Ali
    ,
    Jin, Xin
    ,
    Kramer, George C.
    ,
    Hahn, Jin-Oh
    DOI: 10.1115/1.4052312
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: This paper concerns the design and rigorous in silico evaluation of a closed-loop hemorrhage resuscitation algorithm with blood pressure (BP) as controlled variable. A lumped-parameter control design model relating volume resuscitation input to blood volume (BV) and BP responses was developed and experimentally validated. Then, three alternative adaptive control algorithms were developed using the control design model: (i) model reference adaptive control (MRAC) with BP feedback, (ii) composite adaptive control (CAC) with BP feedback, and (iii) CAC with BV and BP feedback. To the best of our knowledge, this is the first work to demonstrate model-based control design for hemorrhage resuscitation with readily available BP as feedback. The efficacy of these closed-loop control algorithms was comparatively evaluated as well as compared with an empiric expert knowledge-based algorithm based on 100 realistic virtual patients created using a well-established physiological model of cardiovascular (CV) hemodynamics. The in silico evaluation results suggested that the adaptive control algorithms outperformed the knowledge-based algorithm in terms of both accuracy and robustness in BP set point tracking: the average median performance error (MDPE) and median absolute performance error (MDAPE) were significantly smaller by >
     
    99% and >
     
    91%, and as well, their interindividual variability was significantly smaller by >
     
    88% and >
     
    94%. Pending in vivo evaluation, model-based control design may advance the medical autonomy in closed-loop hemorrhage resuscitation.
     
    • Download: (1.635Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Design and In Silico Evaluation of a Closed-Loop Hemorrhage Resuscitation Algorithm With Blood Pressure as Controlled Variable

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4284669
    Collections
    • Journal of Dynamic Systems, Measurement, and Control

    Show full item record

    contributor authorAlsalti, Mohammad
    contributor authorTivay, Ali
    contributor authorJin, Xin
    contributor authorKramer, George C.
    contributor authorHahn, Jin-Oh
    date accessioned2022-05-08T09:02:52Z
    date available2022-05-08T09:02:52Z
    date copyright10/1/2021 12:00:00 AM
    date issued2021
    identifier issn0022-0434
    identifier otherds_144_02_021001.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4284669
    description abstractThis paper concerns the design and rigorous in silico evaluation of a closed-loop hemorrhage resuscitation algorithm with blood pressure (BP) as controlled variable. A lumped-parameter control design model relating volume resuscitation input to blood volume (BV) and BP responses was developed and experimentally validated. Then, three alternative adaptive control algorithms were developed using the control design model: (i) model reference adaptive control (MRAC) with BP feedback, (ii) composite adaptive control (CAC) with BP feedback, and (iii) CAC with BV and BP feedback. To the best of our knowledge, this is the first work to demonstrate model-based control design for hemorrhage resuscitation with readily available BP as feedback. The efficacy of these closed-loop control algorithms was comparatively evaluated as well as compared with an empiric expert knowledge-based algorithm based on 100 realistic virtual patients created using a well-established physiological model of cardiovascular (CV) hemodynamics. The in silico evaluation results suggested that the adaptive control algorithms outperformed the knowledge-based algorithm in terms of both accuracy and robustness in BP set point tracking: the average median performance error (MDPE) and median absolute performance error (MDAPE) were significantly smaller by >
    description abstract99% and >
    description abstract91%, and as well, their interindividual variability was significantly smaller by >
    description abstract88% and >
    description abstract94%. Pending in vivo evaluation, model-based control design may advance the medical autonomy in closed-loop hemorrhage resuscitation.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleDesign and In Silico Evaluation of a Closed-Loop Hemorrhage Resuscitation Algorithm With Blood Pressure as Controlled Variable
    typeJournal Paper
    journal volume144
    journal issue2
    journal titleJournal of Dynamic Systems, Measurement, and Control
    identifier doi10.1115/1.4052312
    journal fristpage21001-1
    journal lastpage21001-10
    page10
    treeJournal of Dynamic Systems, Measurement, and Control:;2021:;volume( 144 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian