contributor author | Irene Doo | |
contributor author | Cz. M. Rodkiewicz | |
date accessioned | 2017-05-08T23:17:21Z | |
date available | 2017-05-08T23:17:21Z | |
date copyright | May, 1984 | |
date issued | 1984 | |
identifier issn | 0148-0731 | |
identifier other | JBENDY-25787#181_1.pdf | |
identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/98173 | |
description abstract | This short note presents the results from experiments performed on an artificial heart valve, in particular, the aortic valve located at the exit of the left ventricle of the heart. The valve was tested in a full-scale elastic model of the arterial system, with a cam-piston unit duplicating the pumping action of the heart. The dependence of the arterial flow distribution on the angular orientation of the valve was investigated and it was found that there was no discernible influence of valve orientation on the arterial flow distribution. | |
publisher | The American Society of Mechanical Engineers (ASME) | |
title | Effect of Aortic Valve Orientation on Arterial Flow Distribution | |
type | Journal Paper | |
journal volume | 106 | |
journal issue | 2 | |
journal title | Journal of Biomechanical Engineering | |
identifier doi | 10.1115/1.3138478 | |
journal fristpage | 181 | |
journal lastpage | 182 | |
identifier eissn | 1528-8951 | |
keywords | Flow (Dynamics) | |
keywords | Valves | |
keywords | Pistons AND Heart valve prostheses | |
tree | Journal of Biomechanical Engineering:;1984:;volume( 106 ):;issue: 002 | |
contenttype | Fulltext | |