contributor authorIrene Doo
contributor authorCz. M. Rodkiewicz
date accessioned2017-05-08T23:17:21Z
date available2017-05-08T23:17:21Z
date copyrightMay, 1984
date issued1984
identifier issn0148-0731
identifier otherJBENDY-25787#181_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/98173
description abstractThis short note presents the results from experiments performed on an artificial heart valve, in particular, the aortic valve located at the exit of the left ventricle of the heart. The valve was tested in a full-scale elastic model of the arterial system, with a cam-piston unit duplicating the pumping action of the heart. The dependence of the arterial flow distribution on the angular orientation of the valve was investigated and it was found that there was no discernible influence of valve orientation on the arterial flow distribution.
publisherThe American Society of Mechanical Engineers (ASME)
titleEffect of Aortic Valve Orientation on Arterial Flow Distribution
typeJournal Paper
journal volume106
journal issue2
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.3138478
journal fristpage181
journal lastpage182
identifier eissn1528-8951
keywordsFlow (Dynamics)
keywordsValves
keywordsPistons AND Heart valve prostheses
treeJournal of Biomechanical Engineering:;1984:;volume( 106 ):;issue: 002
contenttypeFulltext