contributor author | J. C. Ziegert | |
contributor author | J. L. Lewis | |
date accessioned | 2017-05-08T23:06:21Z | |
date available | 2017-05-08T23:06:21Z | |
date copyright | August, 1979 | |
date issued | 1979 | |
identifier issn | 0148-0731 | |
identifier other | JBENDY-25632#218_1.pdf | |
identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/91910 | |
description abstract | The effect of the soft tissue between bone and a preloaded skin surface accelerometer was studied in vivo by comparing its output with the output of an accelerometer connected directly to the bone by a needle through the soft tissue. A 34-g skin surface accelerometer gave an output with little resemblance to the bone motions, appearing to oscillate at its resonant frequency on the soft tissue. A 1.5-g skin surface accelerometer showed nearly identical output to the bone acceleration. | |
publisher | The American Society of Mechanical Engineers (ASME) | |
title | The Effect of Soft Tissue on Measurements of Vibrational Bone Motion by Skin-Mounted Accelerometers | |
type | Journal Paper | |
journal volume | 101 | |
journal issue | 3 | |
journal title | Journal of Biomechanical Engineering | |
identifier doi | 10.1115/1.3426248 | |
journal fristpage | 218 | |
journal lastpage | 220 | |
identifier eissn | 1528-8951 | |
keywords | Measurement | |
keywords | Motion | |
keywords | Accelerometers | |
keywords | Bone | |
keywords | Skin | |
keywords | Soft tissues AND needles | |
tree | Journal of Biomechanical Engineering:;1979:;volume( 101 ):;issue: 003 | |
contenttype | Fulltext | |