    The Development of a Novel Peristaltic Test Stand “Swallow-Sim” for the Mechanical Evaluation of Esophageal Stents

    Source: Journal of Medical Devices:;2024:;volume( 018 ):;issue: 003::page 31005-1
    Author:
    Berger, Daniel Michael
    ,
    Sehrt, Jan T.
    ,
    Brinkmann, Thomas
    ,
    Henrichs, Niclas Norman
    ,
    Bilec, Oguzhan
    ,
    Fuchs, Karl-Hermann
    ,
    Meining, Alexander
    DOI: 10.1115/1.4065906
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: The development of medical products begins with the “in silico” phase, where the development and simulation of new stent types are carried out. This is followed by the “in vitro” phase. Here, tests are done in a test stand to obtain initial conclusions about the interaction of the environment. The approval process is completed in the “in vivo” phase, where testing in living beings happen. Here, preclinical studies are carried out in animals first, followed by clinical studies on patients. A big part of the development fails in this final phase, as this is where the interactions of all influences from the stent environment are investigated. Since this not only causes high costs for the developers but also unnecessarily destroys living resources in animal studies, this publication describes the development of a test stand called “Swallow-Sim” that superimposes the mechanical influences of the esophagus, the chemical stress caused by hydrochloric acid and increased body temperature. Furthermore, tests of the acting pressures are carried out using esophageal manometry, a temperature test of the test stand and a test run of the gastric juice. At the end of this publication, the results are evaluated with a six-week test of a Nickel Titanium Naval Ordnance Laboratory stent, which loses much of its mechanical properties and is partially destroyed by the load. The results show a clear correlation with the findings from reality. The test stand should be further optimized and examined in more detail in further tests and subjected to a reality check.
    • Download: (2.854Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      The Development of a Novel Peristaltic Test Stand “Swallow-Sim” for the Mechanical Evaluation of Esophageal Stents

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4303575
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorBerger, Daniel Michael
    contributor authorSehrt, Jan T.
    contributor authorBrinkmann, Thomas
    contributor authorHenrichs, Niclas Norman
    contributor authorBilec, Oguzhan
    contributor authorFuchs, Karl-Hermann
    contributor authorMeining, Alexander
    date accessioned2024-12-24T19:14:48Z
    date available2024-12-24T19:14:48Z
    date copyright7/30/2024 12:00:00 AM
    date issued2024
    identifier issn1932-6181
    identifier othermed_018_03_031005.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4303575
    description abstractThe development of medical products begins with the “in silico” phase, where the development and simulation of new stent types are carried out. This is followed by the “in vitro” phase. Here, tests are done in a test stand to obtain initial conclusions about the interaction of the environment. The approval process is completed in the “in vivo” phase, where testing in living beings happen. Here, preclinical studies are carried out in animals first, followed by clinical studies on patients. A big part of the development fails in this final phase, as this is where the interactions of all influences from the stent environment are investigated. Since this not only causes high costs for the developers but also unnecessarily destroys living resources in animal studies, this publication describes the development of a test stand called “Swallow-Sim” that superimposes the mechanical influences of the esophagus, the chemical stress caused by hydrochloric acid and increased body temperature. Furthermore, tests of the acting pressures are carried out using esophageal manometry, a temperature test of the test stand and a test run of the gastric juice. At the end of this publication, the results are evaluated with a six-week test of a Nickel Titanium Naval Ordnance Laboratory stent, which loses much of its mechanical properties and is partially destroyed by the load. The results show a clear correlation with the findings from reality. The test stand should be further optimized and examined in more detail in further tests and subjected to a reality check.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleThe Development of a Novel Peristaltic Test Stand “Swallow-Sim” for the Mechanical Evaluation of Esophageal Stents
    typeJournal Paper
    journal volume18
    journal issue3
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4065906
    journal fristpage31005-1
    journal lastpage31005-10
    page10
    treeJournal of Medical Devices:;2024:;volume( 018 ):;issue: 003
    contenttypeFulltext