A Proof of Concept Study of the Mechanical Behavior of Lattice Structures Used to Design a Shoulder Hemi-ProsthesisSource: Journal of Engineering and Science in Medical Diagnostics and Therapy:;2021:;volume( 004 ):;issue: 003::page 031004-1Author:Peto, Marinela
,
Aguilar-Rosas, Oscar
,
Ramirez-Cedillo, Erick
,
Jimenez, Moises
,
Hernandez, Adriana
,
Siller, Hector R.
DOI: 10.1115/1.4051419Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: Lattice structures offer great benefits when employed in medical implants for cell attachment and growth (osseointegration), minimization of stress shielding phenomena, and weight reduction. This study is focused on a proof of concept for developing a generic shoulder hemi-prosthesis, from a patient-specific case of a 46-years-old male with a tumor on the upper part of his humerus. A personalized biomodel was designed and a lattice structure was integrated in its middle portion, to lighten weight without affecting humerus' mechanical response. To select the most appropriate lattice structure, three different configurations were initially tested: tetrahedral vertex centroid (TVC), hexagonal prism vertex centroid (HPVC), and cubic diamond (CD). They were fabricated in resin by digital light processing and its mechanical behavior was studied via compression testing and finite element modeling (FEM). The selected structure according to the results was the HPVC, which was integrated in a digital twin of the biomodel to validate its mechanical performance through FEM but substituting the bone material model with a biocompatible titanium alloy (Ti6Al4V) suitable for prostheses fabrication. Results of the simulation showed acceptable levels of Von Mises stresses (325 MPa max.), below the elastic limit of the titanium alloys, and a better response (52 MPa max.) in a model with equivalent elastic properties, with stress performance in the same order of magnitude than the showed in bone's material model.
|
Show full item record
contributor author | Peto, Marinela | |
contributor author | Aguilar-Rosas, Oscar | |
contributor author | Ramirez-Cedillo, Erick | |
contributor author | Jimenez, Moises | |
contributor author | Hernandez, Adriana | |
contributor author | Siller, Hector R. | |
date accessioned | 2022-02-06T05:40:52Z | |
date available | 2022-02-06T05:40:52Z | |
date copyright | 7/8/2021 12:00:00 AM | |
date issued | 2021 | |
identifier issn | 2572-7958 | |
identifier other | jesmdt_004_03_031004.pdf | |
identifier uri | http://yetl.yabesh.ir/yetl1/handle/yetl/4278534 | |
description abstract | Lattice structures offer great benefits when employed in medical implants for cell attachment and growth (osseointegration), minimization of stress shielding phenomena, and weight reduction. This study is focused on a proof of concept for developing a generic shoulder hemi-prosthesis, from a patient-specific case of a 46-years-old male with a tumor on the upper part of his humerus. A personalized biomodel was designed and a lattice structure was integrated in its middle portion, to lighten weight without affecting humerus' mechanical response. To select the most appropriate lattice structure, three different configurations were initially tested: tetrahedral vertex centroid (TVC), hexagonal prism vertex centroid (HPVC), and cubic diamond (CD). They were fabricated in resin by digital light processing and its mechanical behavior was studied via compression testing and finite element modeling (FEM). The selected structure according to the results was the HPVC, which was integrated in a digital twin of the biomodel to validate its mechanical performance through FEM but substituting the bone material model with a biocompatible titanium alloy (Ti6Al4V) suitable for prostheses fabrication. Results of the simulation showed acceptable levels of Von Mises stresses (325 MPa max.), below the elastic limit of the titanium alloys, and a better response (52 MPa max.) in a model with equivalent elastic properties, with stress performance in the same order of magnitude than the showed in bone's material model. | |
publisher | The American Society of Mechanical Engineers (ASME) | |
title | A Proof of Concept Study of the Mechanical Behavior of Lattice Structures Used to Design a Shoulder Hemi-Prosthesis | |
type | Journal Paper | |
journal volume | 4 | |
journal issue | 3 | |
journal title | Journal of Engineering and Science in Medical Diagnostics and Therapy | |
identifier doi | 10.1115/1.4051419 | |
journal fristpage | 031004-1 | |
journal lastpage | 031004-11 | |
page | 11 | |
tree | Journal of Engineering and Science in Medical Diagnostics and Therapy:;2021:;volume( 004 ):;issue: 003 | |
contenttype | Fulltext |