A Four Degree of Freedom Robot for Positioning Ultrasound Imaging CathetersSource: Journal of Mechanisms and Robotics:;2016:;volume( 008 ):;issue: 005::page 51016Author:Loschak, Paul M.
,
Degirmenci, Alperen
,
Tenzer, Yaroslav
,
Tschabrunn, Cory M.
,
Anter, Elad
,
Howe, Robert D.
DOI: 10.1115/1.4032249Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: In this paper, we present the design, fabrication, and testing of a robot for automatically positioning ultrasound (US) imaging catheters. Our system will point US catheters to provide realtime imaging of anatomical structures and working instruments during minimally invasive procedures. Manually navigating US catheters is difficult and requires extensive training in order to aim the US imager at desired targets. Therefore, a fourdegreeoffreedom (4DOF) robotic system was developed to automatically navigate US imaging catheters for enhanced imaging. A rotational transmission enables 3DOF for pitch, yaw, and roll of the imager. This transmission is translated by the 4DOF. An accuracy analysis calculated the maximum allowable joint motion error. Rotational joints must be accurate to within 1.5 deg, and the translational joint must be accurate within 1.4 mm. Motion tests then validated the accuracy of the robot. The average resulting errors in positioning of the rotational joints were 0.04–0.22 deg. The average measured backlash was 0.18–0.86 deg. Measurements of average translational positioning and backlash errors were negligible. The resulting joint motion errors were well within the required specifications for accurate robot motion. The output of the catheter was then tested to verify the effectiveness of the handle motions to transmit torques and translations to the catheter tip. The catheter tip was navigated to desired target poses with average error 1.3 mm and 0.71 deg. Such effective manipulation of US imaging catheters will enable better visualization in various procedures ranging from cardiac arrhythmia treatment to tumor removal in urological cases.
|
Collections
Show full item record
| contributor author | Loschak, Paul M. | |
| contributor author | Degirmenci, Alperen | |
| contributor author | Tenzer, Yaroslav | |
| contributor author | Tschabrunn, Cory M. | |
| contributor author | Anter, Elad | |
| contributor author | Howe, Robert D. | |
| date accessioned | 2017-05-09T01:31:28Z | |
| date available | 2017-05-09T01:31:28Z | |
| date issued | 2016 | |
| identifier issn | 1942-4302 | |
| identifier other | jmr_008_05_051016.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/161927 | |
| description abstract | In this paper, we present the design, fabrication, and testing of a robot for automatically positioning ultrasound (US) imaging catheters. Our system will point US catheters to provide realtime imaging of anatomical structures and working instruments during minimally invasive procedures. Manually navigating US catheters is difficult and requires extensive training in order to aim the US imager at desired targets. Therefore, a fourdegreeoffreedom (4DOF) robotic system was developed to automatically navigate US imaging catheters for enhanced imaging. A rotational transmission enables 3DOF for pitch, yaw, and roll of the imager. This transmission is translated by the 4DOF. An accuracy analysis calculated the maximum allowable joint motion error. Rotational joints must be accurate to within 1.5 deg, and the translational joint must be accurate within 1.4 mm. Motion tests then validated the accuracy of the robot. The average resulting errors in positioning of the rotational joints were 0.04–0.22 deg. The average measured backlash was 0.18–0.86 deg. Measurements of average translational positioning and backlash errors were negligible. The resulting joint motion errors were well within the required specifications for accurate robot motion. The output of the catheter was then tested to verify the effectiveness of the handle motions to transmit torques and translations to the catheter tip. The catheter tip was navigated to desired target poses with average error 1.3 mm and 0.71 deg. Such effective manipulation of US imaging catheters will enable better visualization in various procedures ranging from cardiac arrhythmia treatment to tumor removal in urological cases. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | A Four Degree of Freedom Robot for Positioning Ultrasound Imaging Catheters | |
| type | Journal Paper | |
| journal volume | 8 | |
| journal issue | 5 | |
| journal title | Journal of Mechanisms and Robotics | |
| identifier doi | 10.1115/1.4032249 | |
| journal fristpage | 51016 | |
| journal lastpage | 51016 | |
| identifier eissn | 1942-4310 | |
| tree | Journal of Mechanisms and Robotics:;2016:;volume( 008 ):;issue: 005 | |
| contenttype | Fulltext |