description abstractA new concept for the evaluation of flowmeter dynamic response is presented. A technique based on hydraulic line dynamics is applied to achieve accurate measurements of the reference flowrate. The major components in the equipment used in experimental verification of the theory were a test pipe, a pulsating flow generator, three pressure transducers, an A/D converter, and a microcomputer system. This equipment allowed the dynamic characteristics of a flowmeter to be investigated in both the time and frequency domains. Comparisons of flowrates measured using the line dynamic technique and the same flowrates recorded by the flowmeter are presented. The results confirm the feasibility and practicability of the dynamic line technique for measuring rapidly varying flows.
publisherThe American Society of Mechanical Engineers (ASME)
titleA Dynamic Evaluation Technique for Liquid Flow Meters Based on Hydraulic Line Dynamics
typeJournal Paper
journal volume119
journal issue2
journal titleJournal of Dynamic Systems, Measurement, and Control
identifier doi10.1115/1.2801252
journal fristpage305
journal lastpage307
identifier eissn1528-9028
keywordsFlowmeters
keywordsDynamics (Mechanics)
keywordsFlow (Dynamics)
keywordsMeasurement
keywordsPressure transducers
keywordsPipes
keywordsDynamic response
keywordsGenerators AND Pulsatile flow
treeJournal of Dynamic Systems, Measurement, and Control:;1997:;volume( 119 ):;issue: 002