Automatic Fungal Glucan Analyzer: For the Rapid Laboratory TestingSource: Journal of Medical Devices:;2025:;volume( 019 ):;issue: 003::page 31001-1DOI: 10.1115/1.4068327Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: Invasive fungal infections (IFIs) have a high mortality rate, especially in the context of invasive treatments such as the use of high-intensity immunosuppressive drugs and indwelling catheters. It is crucial to have a sensitive instrument for rapid laboratory testing to detect early invasive fungal disease (IFD). Therefore, we propose an automatic fungal dextran analyzer that streamlines sample addition, mixing, and result processing, reducing cross-contamination and enhancing operational efficiency. This analyzer utilizes human serum as the sample and offers rapid point-of-care testing (POCT) for IFD without the need for manual intervention. Each individual sample can be processed in just 45 min, and the instrument has the capacity to handle 24 patient specimens within an hour. To address optoelectronic noise concerns, our instrument's algorithm design incorporates both sliding average filtering and a modified version of the least squares method. The application of these techniques significantly enhances the algorithm's accuracy, resulting in an impressive improvement from 95.82% to 99.98%. The instrument underwent clinical testing using serum samples obtained from 130 subjects. The study findings revealed that the instrument effectively facilitated early diagnosis of IFD, thereby considerably improving the efficiency of the testing process.
|
Collections
Show full item record
contributor author | Wu, Yamei | |
contributor author | Liu, Zeyuan | |
contributor author | Zhang, Yu | |
contributor author | Liu, Chunlong | |
contributor author | Zhu, Guimei | |
contributor author | Zhong, Chen | |
date accessioned | 2025-08-20T09:20:21Z | |
date available | 2025-08-20T09:20:21Z | |
date copyright | 5/5/2025 12:00:00 AM | |
date issued | 2025 | |
identifier issn | 1932-6181 | |
identifier other | med_019_03_031001.pdf | |
identifier uri | http://yetl.yabesh.ir/yetl1/handle/yetl/4308112 | |
description abstract | Invasive fungal infections (IFIs) have a high mortality rate, especially in the context of invasive treatments such as the use of high-intensity immunosuppressive drugs and indwelling catheters. It is crucial to have a sensitive instrument for rapid laboratory testing to detect early invasive fungal disease (IFD). Therefore, we propose an automatic fungal dextran analyzer that streamlines sample addition, mixing, and result processing, reducing cross-contamination and enhancing operational efficiency. This analyzer utilizes human serum as the sample and offers rapid point-of-care testing (POCT) for IFD without the need for manual intervention. Each individual sample can be processed in just 45 min, and the instrument has the capacity to handle 24 patient specimens within an hour. To address optoelectronic noise concerns, our instrument's algorithm design incorporates both sliding average filtering and a modified version of the least squares method. The application of these techniques significantly enhances the algorithm's accuracy, resulting in an impressive improvement from 95.82% to 99.98%. The instrument underwent clinical testing using serum samples obtained from 130 subjects. The study findings revealed that the instrument effectively facilitated early diagnosis of IFD, thereby considerably improving the efficiency of the testing process. | |
publisher | The American Society of Mechanical Engineers (ASME) | |
title | Automatic Fungal Glucan Analyzer: For the Rapid Laboratory Testing | |
type | Journal Paper | |
journal volume | 19 | |
journal issue | 3 | |
journal title | Journal of Medical Devices | |
identifier doi | 10.1115/1.4068327 | |
journal fristpage | 31001-1 | |
journal lastpage | 31001-10 | |
page | 10 | |
tree | Journal of Medical Devices:;2025:;volume( 019 ):;issue: 003 | |
contenttype | Fulltext |