Comparative Efficacy of Humidifiers for Noninvasive Infant Respiratory SupportSource: Journal of Medical Devices:;2023:;volume( 017 ):;issue: 001::page 11010-1DOI: 10.1115/1.4056760Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: Delivery of cold, dry air to infants while supporting their breathing can lead to hypothermia in addition to irritating and damaging their sensitive nares and negatively impacting outcomes with these therapies. In high resource settings, electric heated humidifiers are used to mitigate this problem. In many resource-constrained settings, passive nonelectric bubbling humidifiers are instead used. We here compare the efficacy of conventional electric heated humidification, custom-built low cost heated humidification, passive nonelectric bubbling humidification and a control of no humidification. In a hospital patient room (temperature 22 °C, humidity 50%), the temperature and humidity delivered to a simulated patient lung via a BC161-10 Fisher Paykel bubble continuous positive airway pressure (CPAP) system were measured with conventional electric heated humidification, low cost custom-built heated humidification, passive bubbling humidification and no humidification. (Delivered CPAP: 5 cm H2O; flowrate varied from 4 to 8 liters per minute (LPM) in 2 LPM increments.) As the flowrate was varied from 4 to 8 LPM, delivered relative humidity (standard deviation) with each humidifier was as follows: control 10% (3.6%), passive bubbler 44% (3.7%), custom-built humidifier 67% (1.7%), electric heated humidifier 91% (0.86%). Delivered temperature with the electric heated humidifier was 38 °C (0.21 °C) versus 33 °C for all other setups. Conventional electric heating humidification is more effective than passive bubbling humidification, and the custom-built low cost humidifier provides an intermediate degree of humidification. Through further improvement of this concept with a heated inspiratory circuit and sensor based control of the heating element, an effective yet low cost solution heating humidification could be developed.
|
Collections
Show full item record
contributor author | John, Stephen C. | |
contributor author | Hokanson, Casey D. | |
contributor author | Slusher, Tina M. | |
contributor author | Bjorklund, Ashley R. | |
date accessioned | 2023-11-29T19:31:17Z | |
date available | 2023-11-29T19:31:17Z | |
date copyright | 1/31/2023 12:00:00 AM | |
date issued | 1/31/2023 12:00:00 AM | |
date issued | 2023-01-31 | |
identifier issn | 1932-6181 | |
identifier other | med_017_01_011010.pdf | |
identifier uri | http://yetl.yabesh.ir/yetl1/handle/yetl/4294835 | |
description abstract | Delivery of cold, dry air to infants while supporting their breathing can lead to hypothermia in addition to irritating and damaging their sensitive nares and negatively impacting outcomes with these therapies. In high resource settings, electric heated humidifiers are used to mitigate this problem. In many resource-constrained settings, passive nonelectric bubbling humidifiers are instead used. We here compare the efficacy of conventional electric heated humidification, custom-built low cost heated humidification, passive nonelectric bubbling humidification and a control of no humidification. In a hospital patient room (temperature 22 °C, humidity 50%), the temperature and humidity delivered to a simulated patient lung via a BC161-10 Fisher Paykel bubble continuous positive airway pressure (CPAP) system were measured with conventional electric heated humidification, low cost custom-built heated humidification, passive bubbling humidification and no humidification. (Delivered CPAP: 5 cm H2O; flowrate varied from 4 to 8 liters per minute (LPM) in 2 LPM increments.) As the flowrate was varied from 4 to 8 LPM, delivered relative humidity (standard deviation) with each humidifier was as follows: control 10% (3.6%), passive bubbler 44% (3.7%), custom-built humidifier 67% (1.7%), electric heated humidifier 91% (0.86%). Delivered temperature with the electric heated humidifier was 38 °C (0.21 °C) versus 33 °C for all other setups. Conventional electric heating humidification is more effective than passive bubbling humidification, and the custom-built low cost humidifier provides an intermediate degree of humidification. Through further improvement of this concept with a heated inspiratory circuit and sensor based control of the heating element, an effective yet low cost solution heating humidification could be developed. | |
publisher | The American Society of Mechanical Engineers (ASME) | |
title | Comparative Efficacy of Humidifiers for Noninvasive Infant Respiratory Support | |
type | Journal Paper | |
journal volume | 17 | |
journal issue | 1 | |
journal title | Journal of Medical Devices | |
identifier doi | 10.1115/1.4056760 | |
journal fristpage | 11010-1 | |
journal lastpage | 11010-4 | |
page | 4 | |
tree | Journal of Medical Devices:;2023:;volume( 017 ):;issue: 001 | |
contenttype | Fulltext |