    In Silico Mechanical Effort Analysis of the All-On-4 Design Performed With Platform-Switching Distal Short Dental Implants

    Source: Journal of Biomechanical Engineering:;2023:;volume( 145 ):;issue: 009::page 91009-1
    Author:
    Santana, Luís Carlos Leal
    ,
    Idogava, Henrique Takashi
    ,
    De Foggi, Camila Cristina
    ,
    Guastaldi, Fernando Pozzi Semeghini
    ,
    Noritomi, Pedro Yoshito
    ,
    dos Reis, Barbara Araújo
    ,
    Vaz, Luis Geraldo
    DOI: 10.1115/1.4062540
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Short dental implants with platform matching connection have been used for the rehabilitation of atrophic jaws whenever standard-length dental implants cannot be placed without prior bone augmentation. Yet, there remains a lack of data regarding the risk of technical failures when the all-on-4 configuration is performed in atrophic jaws with platform-switching distal short dental implants. Thus, the current study used the finite element method to evaluate the mechanical behavior at the level of the prosthetic components of the all-on-4 concept performed in atrophic mandible using short-length distal implants with platform switching (PSW) connection. Three models of the all-on-4 configuration were generated in human atrophic mandibles. The geometric models consisted of PSW connection tilted standard (AO4T; θ = 30 deg; 11 mm-length), straight standard (AO4S; θ = 0 deg; 11 mm-length) and straight short (AO4Sh; θ = 0 deg; 8 mm-length) distal implants. A resultant force of 300 N was performed obliquely in the left side and posterior region of the prosthetic bar. The von Mises equivalent stress (σvm) and maximum and minimum principal stresses (σmax and σmin) were performed at level of the prosthetic components/implants and peri-implant bone crest, respectively. The general displacement of the models was also evaluated. The stress analysis was performed on the side of load application. The AO4S configuration showed the lowest values of σvm in the mesial left (ML) and distal left (DL) abutments (37.53 MPa and 232.77 MPa, respectively) and dental implants (91.53 MPa and 231.21 MPa, respectively). The AO4Sh configuration showed the highest values of σvm in the bar screw (102.36 MPa), abutment (117.56 MPa), and dental implant (293.73 MPa) of the ML area. Among the models, the highest values of σmax and σmin were noticed in the peri-implant bone crest of the AO4T design (131.48 MPa and 195.31 MPa, respectively). All models showed similar values of general displacements, which were concentrated in the mandible symphysis. The all-on-4 configurations designed with PSW connection and tilted standard (AO4T; θ = 30 deg; 11 mm-length), straight standard (AO4S; θ = 0 deg; 11 mm-length) or straight short (AO4Sh; θ = 0 deg; 8 mm-length) distal implants were not associated with higher odds of technical failures. The AO4Sh design may be a promising option for the prosthetic rehabilitation of atrophic jaws.
    • Download: (2.706Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      In Silico Mechanical Effort Analysis of the All-On-4 Design Performed With Platform-Switching Distal Short Dental Implants

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4294611
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorSantana, Luís Carlos Leal
    contributor authorIdogava, Henrique Takashi
    contributor authorDe Foggi, Camila Cristina
    contributor authorGuastaldi, Fernando Pozzi Semeghini
    contributor authorNoritomi, Pedro Yoshito
    contributor authordos Reis, Barbara Araújo
    contributor authorVaz, Luis Geraldo
    date accessioned2023-11-29T19:08:50Z
    date available2023-11-29T19:08:50Z
    date copyright5/31/2023 12:00:00 AM
    date issued5/31/2023 12:00:00 AM
    date issued2023-05-31
    identifier issn0148-0731
    identifier otherbio_145_09_091009.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4294611
    description abstractShort dental implants with platform matching connection have been used for the rehabilitation of atrophic jaws whenever standard-length dental implants cannot be placed without prior bone augmentation. Yet, there remains a lack of data regarding the risk of technical failures when the all-on-4 configuration is performed in atrophic jaws with platform-switching distal short dental implants. Thus, the current study used the finite element method to evaluate the mechanical behavior at the level of the prosthetic components of the all-on-4 concept performed in atrophic mandible using short-length distal implants with platform switching (PSW) connection. Three models of the all-on-4 configuration were generated in human atrophic mandibles. The geometric models consisted of PSW connection tilted standard (AO4T; θ = 30 deg; 11 mm-length), straight standard (AO4S; θ = 0 deg; 11 mm-length) and straight short (AO4Sh; θ = 0 deg; 8 mm-length) distal implants. A resultant force of 300 N was performed obliquely in the left side and posterior region of the prosthetic bar. The von Mises equivalent stress (σvm) and maximum and minimum principal stresses (σmax and σmin) were performed at level of the prosthetic components/implants and peri-implant bone crest, respectively. The general displacement of the models was also evaluated. The stress analysis was performed on the side of load application. The AO4S configuration showed the lowest values of σvm in the mesial left (ML) and distal left (DL) abutments (37.53 MPa and 232.77 MPa, respectively) and dental implants (91.53 MPa and 231.21 MPa, respectively). The AO4Sh configuration showed the highest values of σvm in the bar screw (102.36 MPa), abutment (117.56 MPa), and dental implant (293.73 MPa) of the ML area. Among the models, the highest values of σmax and σmin were noticed in the peri-implant bone crest of the AO4T design (131.48 MPa and 195.31 MPa, respectively). All models showed similar values of general displacements, which were concentrated in the mandible symphysis. The all-on-4 configurations designed with PSW connection and tilted standard (AO4T; θ = 30 deg; 11 mm-length), straight standard (AO4S; θ = 0 deg; 11 mm-length) or straight short (AO4Sh; θ = 0 deg; 8 mm-length) distal implants were not associated with higher odds of technical failures. The AO4Sh design may be a promising option for the prosthetic rehabilitation of atrophic jaws.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleIn Silico Mechanical Effort Analysis of the All-On-4 Design Performed With Platform-Switching Distal Short Dental Implants
    typeJournal Paper
    journal volume145
    journal issue9
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4062540
    journal fristpage91009-1
    journal lastpage91009-11
    page11
    treeJournal of Biomechanical Engineering:;2023:;volume( 145 ):;issue: 009
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian