Show simple item record

contributor authorChoudhury, Sandeep
contributor authorBanerjee, Anik
contributor authorMajumder, Santanu
contributor authorChowdhury, Amit Roy
date accessioned2023-11-29T19:08:37Z
date available2023-11-29T19:08:37Z
date copyright6/7/2023 12:00:00 AM
date issued6/7/2023 12:00:00 AM
date issued2023-06-07
identifier issn2572-7958
identifier otherjesmdt_006_03_030908.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4294609
description abstractFractures of the maxillofacial region are prevalent in both developed and developing nations. Maxillofacial fractures primarily occur as a result of incidents such as car crashes, physical assaults, and attacks. Although using miniplates to treat maxillofacial fractures is a widely accepted practice, the challenge lies in selecting the appropriate miniplate size that is tailored to the specific dimensions of the fracture in each patient. The study aims to evaluate and determine the most suitable design between a two-hole miniplate and a four-hole miniplate for securing a fractured mandible, with the ultimate goal of creating a customized solution for each patient. A mandible model is chosen with a diagonal length of 82 mm and created perpendicular fracture gaps measuring 1 mm to 3 mm on the left buccolingual section of the solid model. A two-hole and a four-hole miniplate are designed with different thicknesses (ranging from 1 mm to 2 mm) and gap distances (ranging from 6 mm to 12 mm). The miniplates were put together in a model of the mandible. To test their effectiveness, the combination of the mandible model and plates was subjected to various muscle forces, as well as the force of biting, using finite element analysis. At the mandible location, the four-hole miniplate implantation exhibits superior stabilization in comparison to the two-hole miniplate assembly. The miniplate's size and dimensions can be adjusted depending on the size of the fracture in the mandible, resulting in a patient-specific solution for the implantation of miniplate in the mandible.
publisherThe American Society of Mechanical Engineers (ASME)
titleDesign and Development of Patient-Specific Miniplate for the Treatment of Maxillofacial Fractures—A Finite Element Study
typeJournal Paper
journal volume6
journal issue3
journal titleJournal of Engineering and Science in Medical Diagnostics and Therapy
identifier doi10.1115/1.4062509
journal fristpage30908-1
journal lastpage30908-10
page10
treeJournal of Engineering and Science in Medical Diagnostics and Therapy:;2023:;volume( 006 ):;issue: 003
contenttypeFulltext