    Dynamic Analysis of Hip Prosthesis Using Different Biocompatible Alloys

    Source: ASME Open Journal of Engineering:;2022:;volume( 001 )::page 11001-1
    Author:
    Joshi, Tanuj
    ,
    Sharma, Ravikant
    ,
    Mittal, Vinod Kumar
    ,
    Gupta, Vikas
    ,
    Krishan, Gopal
    DOI: 10.1115/1.4053417
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: In the present paper, a three-dimensional finite element model of the Charnley implant has been developed to analyze the stress–strain distribution and deformation over the stem prosthesis. Patient-specific dynamic forces have been considered for the analytical evaluation using commercial finite element code. The impact of each dynamic activity has been analyzed separately using six different biocompatible alloy materials made of titanium and cobalt-chromium. Mechanical parameters have been evaluated to envisage the longevity and functionality of the implant. The performance of different materials for each suitable gait pattern is analyzed using finite element code. Consequently, Cobalt chromium alloys (CoCrMo alloy) demonstrate better results, i.e., maximum stress, minimum deformation, and strain as compared with other materials. Every dynamic motion, viz., walking, standing up, sitting down, going upstairs, and going downstairs are found in good agreement with the safety factor for every biomaterial. Additionally, going downstairs and sitting down gait motion exhibits the maximum and minimum stress–strain level, respectively. Based on the outcome of the presurgical study, it is recommended that CoCrMo alloys should be preferred over other materials.
    • Download: (1.252Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Dynamic Analysis of Hip Prosthesis Using Different Biocompatible Alloys

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4284569
    Collections
    • ASME Open Journal of Engineering

    Show full item record

    contributor authorJoshi, Tanuj
    contributor authorSharma, Ravikant
    contributor authorMittal, Vinod Kumar
    contributor authorGupta, Vikas
    contributor authorKrishan, Gopal
    date accessioned2022-05-08T08:58:10Z
    date available2022-05-08T08:58:10Z
    date copyright2/24/2022 12:00:00 AM
    date issued2022
    identifier issn2770-3495
    identifier otheraoje_1_011001.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4284569
    description abstractIn the present paper, a three-dimensional finite element model of the Charnley implant has been developed to analyze the stress–strain distribution and deformation over the stem prosthesis. Patient-specific dynamic forces have been considered for the analytical evaluation using commercial finite element code. The impact of each dynamic activity has been analyzed separately using six different biocompatible alloy materials made of titanium and cobalt-chromium. Mechanical parameters have been evaluated to envisage the longevity and functionality of the implant. The performance of different materials for each suitable gait pattern is analyzed using finite element code. Consequently, Cobalt chromium alloys (CoCrMo alloy) demonstrate better results, i.e., maximum stress, minimum deformation, and strain as compared with other materials. Every dynamic motion, viz., walking, standing up, sitting down, going upstairs, and going downstairs are found in good agreement with the safety factor for every biomaterial. Additionally, going downstairs and sitting down gait motion exhibits the maximum and minimum stress–strain level, respectively. Based on the outcome of the presurgical study, it is recommended that CoCrMo alloys should be preferred over other materials.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleDynamic Analysis of Hip Prosthesis Using Different Biocompatible Alloys
    typeJournal Paper
    journal volume1
    journal titleASME Open Journal of Engineering
    identifier doi10.1115/1.4053417
    journal fristpage11001-1
    journal lastpage11001-10
    page10
    treeASME Open Journal of Engineering:;2022:;volume( 001 )
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian