    Archive

    Effect of Leaflet Type and Leaflet-Stent Attachment Height on Transcatheter Aortic Valve Leaflet Thrombosis Potential

    Source: Journal of Medical Devices:;2022:;volume( 016 ):;issue: 002::page 21005-1
    Author:
    Ncho, Beatrice E.
    ,
    Siefert, Andrew W.
    ,
    Sadri, Vahid
    ,
    Ortner, Jillian
    ,
    Yoganathan, Ajit P.
    DOI: 10.1115/1.4052902
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Transcatheter aortic valve replacement devices vary in leaflet material and in the height for which leaflets attach to the stented valve frame. Combinations of these features can influence leaflet dynamics, neo-sinus geometries, and fluid dynamics, thereby reducing or exacerbating the potential for blood flow stasis and leaflet thrombosis. To investigate these interconnected relationships, this study evaluated the effects of transcatheter valve leaflet type [porcine versus bovine pericardium] and the leaflet-stent attachment height (low, mid, and high) on flow stasis and potential for leaflet thrombosis. Transcatheter valve models were manufactured and tested within an aortic simulator under pulsatile left heart hemodynamic conditions. Transvalvular hemodynamics, leaflet kinematics, and flow structures were evaluated by direct measurement, high-speed imaging, and two differing techniques of particle image velocimetry. Transcatheter valves with porcine pericardial leaflets were observed to be less stiff, exhibit a lesser resistance to flow, were associated with reduced regions of neo-sinus flow stasis, and superior sinus washout times. More elevated attachments of the leaflets were associated with less neo-sinus flow stasis. These initial results and observations suggest combinations of leaflet type and stent attachment height may reduce transcatheter aortic valve flow stasis and the potential for leaflet thrombosis.
    • Download: (3.786Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Effect of Leaflet Type and Leaflet-Stent Attachment Height on Transcatheter Aortic Valve Leaflet Thrombosis Potential

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4283979
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorNcho, Beatrice E.
    contributor authorSiefert, Andrew W.
    contributor authorSadri, Vahid
    contributor authorOrtner, Jillian
    contributor authorYoganathan, Ajit P.
    date accessioned2022-05-08T08:29:00Z
    date available2022-05-08T08:29:00Z
    date copyright2/3/2022 12:00:00 AM
    date issued2022
    identifier issn1932-6181
    identifier othermed_016_02_021005.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4283979
    description abstractTranscatheter aortic valve replacement devices vary in leaflet material and in the height for which leaflets attach to the stented valve frame. Combinations of these features can influence leaflet dynamics, neo-sinus geometries, and fluid dynamics, thereby reducing or exacerbating the potential for blood flow stasis and leaflet thrombosis. To investigate these interconnected relationships, this study evaluated the effects of transcatheter valve leaflet type [porcine versus bovine pericardium] and the leaflet-stent attachment height (low, mid, and high) on flow stasis and potential for leaflet thrombosis. Transcatheter valve models were manufactured and tested within an aortic simulator under pulsatile left heart hemodynamic conditions. Transvalvular hemodynamics, leaflet kinematics, and flow structures were evaluated by direct measurement, high-speed imaging, and two differing techniques of particle image velocimetry. Transcatheter valves with porcine pericardial leaflets were observed to be less stiff, exhibit a lesser resistance to flow, were associated with reduced regions of neo-sinus flow stasis, and superior sinus washout times. More elevated attachments of the leaflets were associated with less neo-sinus flow stasis. These initial results and observations suggest combinations of leaflet type and stent attachment height may reduce transcatheter aortic valve flow stasis and the potential for leaflet thrombosis.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleEffect of Leaflet Type and Leaflet-Stent Attachment Height on Transcatheter Aortic Valve Leaflet Thrombosis Potential
    typeJournal Paper
    journal volume16
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4052902
    journal fristpage21005-1
    journal lastpage21005-8
    page8
    treeJournal of Medical Devices:;2022:;volume( 016 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian