    A Vibratory, Subresonant Diagnostic Device to Measure Dental Implant Stability Via Angular Stiffness

    Source: Journal of Medical Devices:;2021:;volume( 015 ):;issue: 003::page 031015-1
    Author:
    Xu, Weiwei
    ,
    Wood, Darwin S.
    ,
    Liu, Yifeng
    ,
    Shen, I. Y.
    DOI: 10.1115/1.4051832
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Stability of a dental implant reflects quality of osseointegration between the implant and its surrounding bone. While many methods have been proposed to characterize implant stability, angular stiffness at the neck of the implant has been proven to be a rigorous and accurate measure. Nevertheless, fast and reliable measurements of the angular stiffness in a clinical environment are not yet available. This article is to demonstrate a novel stability diagnostic device that can measure the angular stiffness accurately in clinical environments. The device consists of a sensing unit, a controller unit, and a mobile app. In the sensing unit, a coupler attaches a buzzer motor and a tiny accelerometer to an abutment of an implant, whose angular stiffness is to be measured. The buzzer vibrates at a frequency below the resonance frequency of the implant–bone–abutment system. Meanwhile, the accelerometer measures the abutment's vibration. The controller unit powers the buzzer, reads the accelerometer data, and transmits the data to the mobile app. The mobile app postprocesses the data and extracts the angular stiffness through use of a finite element model and a nonlinear regression algorithm. The extracted angular stiffness is compared with a calibrated angular stiffness, which is obtained independently via a force hammer and a laser Doppler vibrometer. The comparison shows reasonable agreement, with the difference being in the range of 4–20%.
    • Download: (1.343Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      A Vibratory, Subresonant Diagnostic Device to Measure Dental Implant Stability Via Angular Stiffness

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4278737
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorXu, Weiwei
    contributor authorWood, Darwin S.
    contributor authorLiu, Yifeng
    contributor authorShen, I. Y.
    date accessioned2022-02-06T05:46:38Z
    date available2022-02-06T05:46:38Z
    date copyright8/2/2021 12:00:00 AM
    date issued2021
    identifier issn1932-6181
    identifier othermed_015_03_031015.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4278737
    description abstractStability of a dental implant reflects quality of osseointegration between the implant and its surrounding bone. While many methods have been proposed to characterize implant stability, angular stiffness at the neck of the implant has been proven to be a rigorous and accurate measure. Nevertheless, fast and reliable measurements of the angular stiffness in a clinical environment are not yet available. This article is to demonstrate a novel stability diagnostic device that can measure the angular stiffness accurately in clinical environments. The device consists of a sensing unit, a controller unit, and a mobile app. In the sensing unit, a coupler attaches a buzzer motor and a tiny accelerometer to an abutment of an implant, whose angular stiffness is to be measured. The buzzer vibrates at a frequency below the resonance frequency of the implant–bone–abutment system. Meanwhile, the accelerometer measures the abutment's vibration. The controller unit powers the buzzer, reads the accelerometer data, and transmits the data to the mobile app. The mobile app postprocesses the data and extracts the angular stiffness through use of a finite element model and a nonlinear regression algorithm. The extracted angular stiffness is compared with a calibrated angular stiffness, which is obtained independently via a force hammer and a laser Doppler vibrometer. The comparison shows reasonable agreement, with the difference being in the range of 4–20%.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleA Vibratory, Subresonant Diagnostic Device to Measure Dental Implant Stability Via Angular Stiffness
    typeJournal Paper
    journal volume15
    journal issue3
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4051832
    journal fristpage031015-1
    journal lastpage031015-7
    page7
    treeJournal of Medical Devices:;2021:;volume( 015 ):;issue: 003
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian