    An Adaptable Computed Tomography-Derived Three-Dimensional-Printed Alignment Fixture Minimizes Errors in Radius Biomechanical Testing

    Source: Journal of Biomechanical Engineering:;2021:;volume( 143 ):;issue: 011::page 0111006-1
    Author:
    Inacio, Jordan V.
    ,
    Cristino, Danielle M.
    ,
    Hast, Michael W.
    ,
    Dailey, Hannah L.
    DOI: 10.1115/1.4051433
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Biomechanical testing of long bones can be susceptible to errors and uncertainty due to malalignment of specimens with respect to the mechanical axis of the test frame. To solve this problem, we designed a novel, customizable alignment and potting fixture for long bone testing. The fixture consists of three-dimensional-printed components modeled from specimen-specific computed tomography (CT) scans to achieve a predetermined specimen alignment. We demonstrated the functionality of this fixture by comparing benchtop torsional test results to specimen-matched finite element models and found a strong correlation (R2 = 0.95, p < 0.001). Additional computational models were used to estimate the impact of malalignment on mechanical behavior in both torsion and axial compression. Results confirmed that torsion testing is relatively robust to alignment artifacts, with absolute percent errors less than 8% in all malalignment scenarios. In contrast, axial testing was highly sensitive to setup errors, experiencing absolute percent errors up to 50% with off-center malalignment and up to 170% with angular malalignment. This suggests that whenever appropriate, torsion tests should be used preferentially as a summary mechanical measure. When more challenging modes of loading are required, pretest clinical-resolution CT scanning can be effectively used to create potting fixtures that allow for precise preplanned specimen alignment. This may be particularly important for more sensitive biomechanical tests (e.g., axial compressive tests) that may be needed for industrial applications, such as orthopedic implant design.
    • Download: (2.506Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      An Adaptable Computed Tomography-Derived Three-Dimensional-Printed Alignment Fixture Minimizes Errors in Radius Biomechanical Testing

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4278452
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorInacio, Jordan V.
    contributor authorCristino, Danielle M.
    contributor authorHast, Michael W.
    contributor authorDailey, Hannah L.
    date accessioned2022-02-06T05:38:26Z
    date available2022-02-06T05:38:26Z
    date copyright7/19/2021 12:00:00 AM
    date issued2021
    identifier issn0148-0731
    identifier otherbio_143_11_111006.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4278452
    description abstractBiomechanical testing of long bones can be susceptible to errors and uncertainty due to malalignment of specimens with respect to the mechanical axis of the test frame. To solve this problem, we designed a novel, customizable alignment and potting fixture for long bone testing. The fixture consists of three-dimensional-printed components modeled from specimen-specific computed tomography (CT) scans to achieve a predetermined specimen alignment. We demonstrated the functionality of this fixture by comparing benchtop torsional test results to specimen-matched finite element models and found a strong correlation (R2 = 0.95, p < 0.001). Additional computational models were used to estimate the impact of malalignment on mechanical behavior in both torsion and axial compression. Results confirmed that torsion testing is relatively robust to alignment artifacts, with absolute percent errors less than 8% in all malalignment scenarios. In contrast, axial testing was highly sensitive to setup errors, experiencing absolute percent errors up to 50% with off-center malalignment and up to 170% with angular malalignment. This suggests that whenever appropriate, torsion tests should be used preferentially as a summary mechanical measure. When more challenging modes of loading are required, pretest clinical-resolution CT scanning can be effectively used to create potting fixtures that allow for precise preplanned specimen alignment. This may be particularly important for more sensitive biomechanical tests (e.g., axial compressive tests) that may be needed for industrial applications, such as orthopedic implant design.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleAn Adaptable Computed Tomography-Derived Three-Dimensional-Printed Alignment Fixture Minimizes Errors in Radius Biomechanical Testing
    typeJournal Paper
    journal volume143
    journal issue11
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4051433
    journal fristpage0111006-1
    journal lastpage0111006-9
    page9
    treeJournal of Biomechanical Engineering:;2021:;volume( 143 ):;issue: 011
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian