    Improving Endo-Myocardial Biopsy by Real-Time Spectral Tissue Sensing: A Feasibility Study

    Source: Journal of Medical Devices:;2020:;volume( 014 ):;issue: 004::page 041006-1
    Author:
    Ali, Awaz
    ,
    Breedveld, Paul
    ,
    Hendriks, Benno H. W.
    DOI: 10.1115/1.4048374
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: The aim of this research is to integrate spectral tissue sensing technology inside a cardiac bioptome for real-time measurements of tissue characteristics. Bioptome tip and handle components were designed and manufactured to house and guide optical fibers. The designed components were assembled on a cardiac bioptome together with optical fibers. A technical feasibility test was carried out to study the functionality of the instrument and the effect of the optical technology on the biopsy performance. Biopsy samples were taken from five different tissue types in a porcine heart and the resulting optical spectra were compared. Spectral tissue sensing fibers were successfully integrated inside a conventional cardiac bioptome. The integrated instrument allowed differentiation between ventricular tissue, blood, and cardiac fat tissue based on blood and fat percentage and amount of scattering. Moreover, differences between scarred and nonscarred tissue were clearly visible. A first step has been made in the use of spectral tissue sensing for the detection of different tissue structures for endo-myocardial biopsy (EMB). The instrument was able to differentiate between various tissues, as well as between healthy and diseased cardiac tissues. Future research should focus on measurements of naturally diseased cardiac tissue, repeated measurements with statistical value, and improvements to the instrument design. Having the ability to measure tissue characteristics prior to acquiring a biopsy sample will not only allow easier positioning of the bioptome at the correct location but can also prevent sampling undesired tissue or scar tissue from previous biopsies.
    • Download: (1.642Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Improving Endo-Myocardial Biopsy by Real-Time Spectral Tissue Sensing: A Feasibility Study

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4275201
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorAli, Awaz
    contributor authorBreedveld, Paul
    contributor authorHendriks, Benno H. W.
    date accessioned2022-02-04T22:15:27Z
    date available2022-02-04T22:15:27Z
    date copyright10/1/2020 12:00:00 AM
    date issued2020
    identifier issn1932-6181
    identifier othermed_014_04_041006.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4275201
    description abstractThe aim of this research is to integrate spectral tissue sensing technology inside a cardiac bioptome for real-time measurements of tissue characteristics. Bioptome tip and handle components were designed and manufactured to house and guide optical fibers. The designed components were assembled on a cardiac bioptome together with optical fibers. A technical feasibility test was carried out to study the functionality of the instrument and the effect of the optical technology on the biopsy performance. Biopsy samples were taken from five different tissue types in a porcine heart and the resulting optical spectra were compared. Spectral tissue sensing fibers were successfully integrated inside a conventional cardiac bioptome. The integrated instrument allowed differentiation between ventricular tissue, blood, and cardiac fat tissue based on blood and fat percentage and amount of scattering. Moreover, differences between scarred and nonscarred tissue were clearly visible. A first step has been made in the use of spectral tissue sensing for the detection of different tissue structures for endo-myocardial biopsy (EMB). The instrument was able to differentiate between various tissues, as well as between healthy and diseased cardiac tissues. Future research should focus on measurements of naturally diseased cardiac tissue, repeated measurements with statistical value, and improvements to the instrument design. Having the ability to measure tissue characteristics prior to acquiring a biopsy sample will not only allow easier positioning of the bioptome at the correct location but can also prevent sampling undesired tissue or scar tissue from previous biopsies.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleImproving Endo-Myocardial Biopsy by Real-Time Spectral Tissue Sensing: A Feasibility Study
    typeJournal Paper
    journal volume14
    journal issue4
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4048374
    journal fristpage041006-1
    journal lastpage041006-9
    page9
    treeJournal of Medical Devices:;2020:;volume( 014 ):;issue: 004
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian