Biomechanical Analysis to Probe Role of Bone Condition and Subject Weight in Stiffness Customization of Femoral Stem for Improved Periprosthetic Biomechanical ResponseSource: Journal of Biomechanical Engineering:;2020:;volume( 142 ):;issue: 010::page 0101002-1DOI: 10.1115/1.4046973Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: Stress shielding due to difference in stiffness of bone and implant material is one among the foremost causes of loosening and failure of load-bearing implants. Thus far, femoral geometry has been given priority for the customization of total hip joint replacement (THR) implant design. This study, for the first time, demonstrates the key role of bone condition and subject-weight on the customization of stiffness and design of the femoral stem. In particular, internal hollowness was incorporated to reduce the implant stiffness and such designed structure has been customized based on subject parameters, including bone condition and bodyweight. The primary aim was to tailor these parameters to achieve close to natural strain distribution at periprosthetic bone and to reduce interfacial bone loss over time. The maintenance of interfacial bone density over time has been studied here through analysis of bone remodeling (BR). For normal bodyweight, the highest hollowness exhibited clinically relevant biomechanical response, for all bone conditions. However, for heavier subjects, consideration of bone quality was found to be essential as higher hollowness induced bone failure in weaker bones and implant failure in stronger bones. Moreover, for stronger bone, thinner medial wall was found to reduce bone resorption over time on the proximo-lateral zone of stress shielding, while lateral thinning was found advantageous for weaker bones. The findings of this study are likely to facilitate designing of femoral stems for achieving better physiological outcomes and enhancement of the quality of life of patients undergoing THR surgery.
|
Collections
Show full item record
contributor author | Chatterjee, Subhomoy | |
contributor author | Roy, Sandipan | |
contributor author | Majumder, Santanu | |
contributor author | RoyChowdhury, Amit | |
date accessioned | 2022-02-04T21:59:08Z | |
date available | 2022-02-04T21:59:08Z | |
date copyright | 7/8/2020 12:00:00 AM | |
date issued | 2020 | |
identifier issn | 0148-0731 | |
identifier other | bio_142_10_101002.pdf | |
identifier uri | http://yetl.yabesh.ir/yetl1/handle/yetl/4274652 | |
description abstract | Stress shielding due to difference in stiffness of bone and implant material is one among the foremost causes of loosening and failure of load-bearing implants. Thus far, femoral geometry has been given priority for the customization of total hip joint replacement (THR) implant design. This study, for the first time, demonstrates the key role of bone condition and subject-weight on the customization of stiffness and design of the femoral stem. In particular, internal hollowness was incorporated to reduce the implant stiffness and such designed structure has been customized based on subject parameters, including bone condition and bodyweight. The primary aim was to tailor these parameters to achieve close to natural strain distribution at periprosthetic bone and to reduce interfacial bone loss over time. The maintenance of interfacial bone density over time has been studied here through analysis of bone remodeling (BR). For normal bodyweight, the highest hollowness exhibited clinically relevant biomechanical response, for all bone conditions. However, for heavier subjects, consideration of bone quality was found to be essential as higher hollowness induced bone failure in weaker bones and implant failure in stronger bones. Moreover, for stronger bone, thinner medial wall was found to reduce bone resorption over time on the proximo-lateral zone of stress shielding, while lateral thinning was found advantageous for weaker bones. The findings of this study are likely to facilitate designing of femoral stems for achieving better physiological outcomes and enhancement of the quality of life of patients undergoing THR surgery. | |
publisher | The American Society of Mechanical Engineers (ASME) | |
title | Biomechanical Analysis to Probe Role of Bone Condition and Subject Weight in Stiffness Customization of Femoral Stem for Improved Periprosthetic Biomechanical Response | |
type | Journal Paper | |
journal volume | 142 | |
journal issue | 10 | |
journal title | Journal of Biomechanical Engineering | |
identifier doi | 10.1115/1.4046973 | |
journal fristpage | 0101002-1 | |
journal lastpage | 0101002-14 | |
page | 14 | |
tree | Journal of Biomechanical Engineering:;2020:;volume( 142 ):;issue: 010 | |
contenttype | Fulltext |