    Infusion Jet Flow Control in Neonatal Double Lumen Cannulae

    Source: Journal of Biomechanical Engineering:;2020:;volume( 142 ):;issue: 005
    Author:
    Rasooli, Reza
    ,
    Yıldız, Yahya
    ,
    Jamil, Muhammad
    ,
    Pekkan, Kerem
    DOI: 10.1115/1.4044548
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Clinical success of extracorporeal membrane oxygenation (ECMO) depends on the proper venous cannulation. Venovenous (VV) ECMO is the preferred clinical intervention as it provides a single-site access by utilizing a VV double lumen cannula (VVDLC) with a higher level of mobilization and physical rehabilitation. Concurrent venous blood drainage and oxygenated blood infusion in the right atrium at the presence of the cannula makes the flow dynamics complex where potential mixing of venous and oxygenated blood can drastically decrease the overall performance of ECMO. There are no studies focusing on the neonatal and pediatric populations, in which the flow related effects are critical due to the small atrium size. In this study, fluid dynamics of infusion outflow jet for two commercially available neonatal VVDLC is analyzed using particle image velocimetry (PIV). Moreover, six new designs are proposed for the infusion channel geometry and compared. Important flow parameters such as flow turning angle (FTA), velocity decay, potential core, and turbulent intensity are investigated for the proposed models. The experiments showed that the outflow parameters of commercial cannulae such as FTA are strongly dependent on the operating Re number. This may result in a drastic efficiency reduction for cannula operating at off-design flow conditions. Moreover, the infusion outlet tip structure and jet internal guiding pathway (JIGP) was observed to greatly affect the outflow flow features. This is of paramount importance since the anatomical positioning of the cannula and the infusion outlet is strongly dependent on the outflow properties such as FTA.
    • Download: (3.960Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Infusion Jet Flow Control in Neonatal Double Lumen Cannulae

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4274325
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorRasooli, Reza
    contributor authorYıldız, Yahya
    contributor authorJamil, Muhammad
    contributor authorPekkan, Kerem
    date accessioned2022-02-04T14:45:58Z
    date available2022-02-04T14:45:58Z
    date copyright2020/01/17/
    date issued2020
    identifier issn0148-0731
    identifier otherbio_142_05_051002.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4274325
    description abstractClinical success of extracorporeal membrane oxygenation (ECMO) depends on the proper venous cannulation. Venovenous (VV) ECMO is the preferred clinical intervention as it provides a single-site access by utilizing a VV double lumen cannula (VVDLC) with a higher level of mobilization and physical rehabilitation. Concurrent venous blood drainage and oxygenated blood infusion in the right atrium at the presence of the cannula makes the flow dynamics complex where potential mixing of venous and oxygenated blood can drastically decrease the overall performance of ECMO. There are no studies focusing on the neonatal and pediatric populations, in which the flow related effects are critical due to the small atrium size. In this study, fluid dynamics of infusion outflow jet for two commercially available neonatal VVDLC is analyzed using particle image velocimetry (PIV). Moreover, six new designs are proposed for the infusion channel geometry and compared. Important flow parameters such as flow turning angle (FTA), velocity decay, potential core, and turbulent intensity are investigated for the proposed models. The experiments showed that the outflow parameters of commercial cannulae such as FTA are strongly dependent on the operating Re number. This may result in a drastic efficiency reduction for cannula operating at off-design flow conditions. Moreover, the infusion outlet tip structure and jet internal guiding pathway (JIGP) was observed to greatly affect the outflow flow features. This is of paramount importance since the anatomical positioning of the cannula and the infusion outlet is strongly dependent on the outflow properties such as FTA.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleInfusion Jet Flow Control in Neonatal Double Lumen Cannulae
    typeJournal Paper
    journal volume142
    journal issue5
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4044548
    page51002
    treeJournal of Biomechanical Engineering:;2020:;volume( 142 ):;issue: 005
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian