    A Novel Intermediate Attachment to Reduce Contamination in Reusable Core Needle Biopsy Devices

    Source: Journal of Medical Devices:;2020:;volume( 014 ):;issue: 001
    Author:
    Berges, Alexandra J.
    ,
    Callanan, Megan
    ,
    Zawicki, Valerie
    ,
    Shi, Richard
    ,
    Athey, Thomas
    ,
    Ayyappan, Vinay
    ,
    Metzger, Schuyler
    ,
    Farrell, Alanna
    ,
    Manbachi, Amir
    ,
    Harvey, Susan
    ,
    Durr, Nicholas J.
    DOI: 10.1115/1.4045967
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: One barrier to breast cancer diagnosis in low-resource settings is that devices for core needle biopsy (CNB) are either disposable and expensive, or reusable and susceptible to internal contamination. Through interviews with field workers and verification experiments, we identified that a common, commercially available, reusable CNB device allows contaminants to enter the driver chamber during firing, necessitating laborious cleaning of the entire device after every use. We introduce a novel CNB device attachment that eliminates this contamination mode and interfaces with existing commercial reusable drivers and low-cost disposable needles. This attachment repositions the driver–needle connection to the exterior of the driver, preventing retrograde flow of blood. Using an unmodified commercial CNB, we replicate chamber contamination by firing into a body fluid-mimicking glycerol solution. Prototypes were tested for their performance in eliminating this contamination. We tested the effectiveness of a cleaning procedure to reduce trace contamination by using a fluorescent dye and measuring the intensity of fluorescence after cleaning. The device's ability to reliably and consistently biopsy tissue with the novel attachment was evaluated using breast tissue models. In these tests, a reusable CNB with our attachment exhibited no measurable internal contamination, and maintained full biopsy functionality as measured by tissue sample weight and length. Minimizing internal device contamination would simplify the cleaning process for reusable biopsy devices. This would improve the accessibility of breast cancer biopsies in low- and middle-income countries (LMICs).
    • Download: (939.0Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      A Novel Intermediate Attachment to Reduce Contamination in Reusable Core Needle Biopsy Devices

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4274262
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorBerges, Alexandra J.
    contributor authorCallanan, Megan
    contributor authorZawicki, Valerie
    contributor authorShi, Richard
    contributor authorAthey, Thomas
    contributor authorAyyappan, Vinay
    contributor authorMetzger, Schuyler
    contributor authorFarrell, Alanna
    contributor authorManbachi, Amir
    contributor authorHarvey, Susan
    contributor authorDurr, Nicholas J.
    date accessioned2022-02-04T14:44:07Z
    date available2022-02-04T14:44:07Z
    date copyright2020/02/05/
    date issued2020
    identifier issn1932-6181
    identifier othermed_014_01_011107.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4274262
    description abstractOne barrier to breast cancer diagnosis in low-resource settings is that devices for core needle biopsy (CNB) are either disposable and expensive, or reusable and susceptible to internal contamination. Through interviews with field workers and verification experiments, we identified that a common, commercially available, reusable CNB device allows contaminants to enter the driver chamber during firing, necessitating laborious cleaning of the entire device after every use. We introduce a novel CNB device attachment that eliminates this contamination mode and interfaces with existing commercial reusable drivers and low-cost disposable needles. This attachment repositions the driver–needle connection to the exterior of the driver, preventing retrograde flow of blood. Using an unmodified commercial CNB, we replicate chamber contamination by firing into a body fluid-mimicking glycerol solution. Prototypes were tested for their performance in eliminating this contamination. We tested the effectiveness of a cleaning procedure to reduce trace contamination by using a fluorescent dye and measuring the intensity of fluorescence after cleaning. The device's ability to reliably and consistently biopsy tissue with the novel attachment was evaluated using breast tissue models. In these tests, a reusable CNB with our attachment exhibited no measurable internal contamination, and maintained full biopsy functionality as measured by tissue sample weight and length. Minimizing internal device contamination would simplify the cleaning process for reusable biopsy devices. This would improve the accessibility of breast cancer biopsies in low- and middle-income countries (LMICs).
    publisherThe American Society of Mechanical Engineers (ASME)
    titleA Novel Intermediate Attachment to Reduce Contamination in Reusable Core Needle Biopsy Devices
    typeJournal Paper
    journal volume14
    journal issue1
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4045967
    page11107
    treeJournal of Medical Devices:;2020:;volume( 014 ):;issue: 001
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian