    Design and Biomechanical Analysis of a Novel Retrievable Peripheral Vascular Stent

    Source: Journal of Medical Devices:;2020:;volume( 014 ):;issue: 002
    Author:
    Guo, Jingzhen
    ,
    Mao, Lin
    ,
    Yu, Xitong
    ,
    Song, Chengli
    ,
    Zhao, Xue
    DOI: 10.1115/1.4046796
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Structurally retrievable drug-eluting stents may have valuable clinical applications because they do not leave any foreign materials inside the patient's body. This article presents a novel design of retrievable peripheral vascular stent and the results from biomechanical analysis of its performance. Using the finite element analysis method, principal parameters of the stent were studied. Moreover, to ensure the practicability of the retrieval process, simulation, and in vitro experiments were performed. The retrieval force reached the maximum value when the whole retrievable part had been retrieved. Furthermore, the force was gradually increased during the retrieval process and remained constant after the main part had been retrieved. When the stent was being compressed, the maximum strain of the stent occurred at the connection between the stent's retrieval part and the main body part, at a value of 4%. The index of nonuniformity of the stent was too small to be counted both at the end of the compression and self-expansion processes. With the increase of moment, the bending stiffness (EI) of the stent decreased gradually. After bending moment was applied, the large strain region was mainly located in the stent's main body part rather than the retrieval part. The results of preliminary stent retrieval experiments demonstrated that the stent could be retrieved successfully. This novel retrievable stent displays promising biomechanical performance. The preliminary experiments demonstrated that the stent could be retrieved smoothly from the blood vessels.
    • Download: (2.291Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Design and Biomechanical Analysis of a Novel Retrievable Peripheral Vascular Stent

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4273741
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorGuo, Jingzhen
    contributor authorMao, Lin
    contributor authorYu, Xitong
    contributor authorSong, Chengli
    contributor authorZhao, Xue
    date accessioned2022-02-04T14:28:54Z
    date available2022-02-04T14:28:54Z
    date copyright2020/04/10/
    date issued2020
    identifier issn1932-6181
    identifier othermed_014_02_025006.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4273741
    description abstractStructurally retrievable drug-eluting stents may have valuable clinical applications because they do not leave any foreign materials inside the patient's body. This article presents a novel design of retrievable peripheral vascular stent and the results from biomechanical analysis of its performance. Using the finite element analysis method, principal parameters of the stent were studied. Moreover, to ensure the practicability of the retrieval process, simulation, and in vitro experiments were performed. The retrieval force reached the maximum value when the whole retrievable part had been retrieved. Furthermore, the force was gradually increased during the retrieval process and remained constant after the main part had been retrieved. When the stent was being compressed, the maximum strain of the stent occurred at the connection between the stent's retrieval part and the main body part, at a value of 4%. The index of nonuniformity of the stent was too small to be counted both at the end of the compression and self-expansion processes. With the increase of moment, the bending stiffness (EI) of the stent decreased gradually. After bending moment was applied, the large strain region was mainly located in the stent's main body part rather than the retrieval part. The results of preliminary stent retrieval experiments demonstrated that the stent could be retrieved successfully. This novel retrievable stent displays promising biomechanical performance. The preliminary experiments demonstrated that the stent could be retrieved smoothly from the blood vessels.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleDesign and Biomechanical Analysis of a Novel Retrievable Peripheral Vascular Stent
    typeJournal Paper
    journal volume14
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4046796
    page25006
    treeJournal of Medical Devices:;2020:;volume( 014 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian