    Development and Assessment of a Microcomputed Tomography Compatible Five Degrees-of-Freedom Knee Joint Motion Simulator

    Source: Journal of Biomechanical Engineering:;2019:;volume( 141 ):;issue: 010::page 101006
    Author:
    Blokker, Alexandra M.
    ,
    Getgood, Alan M.
    ,
    Curiale, Nathan J.
    ,
    Nikolov, Hristo N.
    ,
    Laing, Justin G.
    ,
    Holdsworth, David W.
    ,
    Burkhart, Timothy A.
    DOI: 10.1115/1.4043755
    Publisher: American Society of Mechanical Engineers (ASME)
    Abstract: Currently available knee joint kinematic tracking systems fail to nondestructively capture the subtle variation in joint and soft tissue kinematics that occur in native, injured, and reconstructed joint states. Microcomputed tomography (CT) imaging has the potential as a noninvasive, high-resolution kinematic tracking system, but no dynamic simulators exist to take advantage of this. The purpose of this work was to develop and assess a novel micro-CT compatible knee joint simulator to quantify the knee joint's kinematic and kinetic response to clinically (e.g., pivot shift test) and functionally (e.g., gait) relevant loading. The simulator applies closed-loop, load control over four degrees-of-freedom (DOF) (internal/external rotation, varus/valgus rotation, anterior/posterior translation, and compression/distraction), and static control over a fifth degree-of-freedom (flexion/extension). Simulator accuracy (e.g., load error) and repeatability (e.g., coefficient of variation) were assessed with a cylindrical rubber tubing structure and a human cadaveric knee joint by applying clinically and functionally relevant loads along all active axes. Micro-CT images acquired of the joint at a loaded state were then used to calculate joint kinematics. The simulator loaded both the rubber tubing and the cadaveric specimen to within 0.1% of the load target, with an intertrial coefficient of variation below 0.1% for all clinically relevant loading protocols. The resultant kinematics calculated from the acquired images agreed with previously published values, and produced errors of 1.66 mm, 0.90 mm, 4.41 deg, and 1.60 deg with respect to anterior translation, compression, internal rotation, and valgus rotation, respectively. All images were free of artifacts and showed knee joint displacements in response to clinically and functionally loading with isotropic CT image voxel spacing of 0.15 mm. The results of this study demonstrate that the joint-motion simulator is capable of applying accurate, clinically and functionally relevant loads to cadaveric knee joints, concurrent with micro-CT imaging. Nondestructive tracking of bony landmarks allows for the precise calculation of joint kinematics with less error than traditional optical tracking systems.
    • Download: (2.422Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Development and Assessment of a Microcomputed Tomography Compatible Five Degrees-of-Freedom Knee Joint Motion Simulator

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4258054
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorBlokker, Alexandra M.
    contributor authorGetgood, Alan M.
    contributor authorCuriale, Nathan J.
    contributor authorNikolov, Hristo N.
    contributor authorLaing, Justin G.
    contributor authorHoldsworth, David W.
    contributor authorBurkhart, Timothy A.
    date accessioned2019-09-18T09:01:52Z
    date available2019-09-18T09:01:52Z
    date copyright7/15/2019 12:00:00 AM
    date issued2019
    identifier issn0148-0731
    identifier otherbio_141_10_101006
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4258054
    description abstractCurrently available knee joint kinematic tracking systems fail to nondestructively capture the subtle variation in joint and soft tissue kinematics that occur in native, injured, and reconstructed joint states. Microcomputed tomography (CT) imaging has the potential as a noninvasive, high-resolution kinematic tracking system, but no dynamic simulators exist to take advantage of this. The purpose of this work was to develop and assess a novel micro-CT compatible knee joint simulator to quantify the knee joint's kinematic and kinetic response to clinically (e.g., pivot shift test) and functionally (e.g., gait) relevant loading. The simulator applies closed-loop, load control over four degrees-of-freedom (DOF) (internal/external rotation, varus/valgus rotation, anterior/posterior translation, and compression/distraction), and static control over a fifth degree-of-freedom (flexion/extension). Simulator accuracy (e.g., load error) and repeatability (e.g., coefficient of variation) were assessed with a cylindrical rubber tubing structure and a human cadaveric knee joint by applying clinically and functionally relevant loads along all active axes. Micro-CT images acquired of the joint at a loaded state were then used to calculate joint kinematics. The simulator loaded both the rubber tubing and the cadaveric specimen to within 0.1% of the load target, with an intertrial coefficient of variation below 0.1% for all clinically relevant loading protocols. The resultant kinematics calculated from the acquired images agreed with previously published values, and produced errors of 1.66 mm, 0.90 mm, 4.41 deg, and 1.60 deg with respect to anterior translation, compression, internal rotation, and valgus rotation, respectively. All images were free of artifacts and showed knee joint displacements in response to clinically and functionally loading with isotropic CT image voxel spacing of 0.15 mm. The results of this study demonstrate that the joint-motion simulator is capable of applying accurate, clinically and functionally relevant loads to cadaveric knee joints, concurrent with micro-CT imaging. Nondestructive tracking of bony landmarks allows for the precise calculation of joint kinematics with less error than traditional optical tracking systems.
    publisherAmerican Society of Mechanical Engineers (ASME)
    titleDevelopment and Assessment of a Microcomputed Tomography Compatible Five Degrees-of-Freedom Knee Joint Motion Simulator
    typeJournal Paper
    journal volume141
    journal issue10
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4043755
    journal fristpage101006
    journal lastpage101006-10
    treeJournal of Biomechanical Engineering:;2019:;volume( 141 ):;issue: 010
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian