Show simple item record

contributor authorTeo, Tabitha H. T.
contributor authorRamani, Ajay
contributor authorMunden, Paul M.
contributor authorWilson, Sara E.
contributor authorKieweg, Sarah L.
contributor authorDougherty, Ronald L.
date accessioned2019-03-17T10:24:24Z
date available2019-03-17T10:24:24Z
date copyright7/3/2018 12:00:00 AM
date issued2019
identifier issn2572-7958
identifier otherjesmdt_001_04_041001.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4256123
description abstractTreatment of vision-threating elevated intraocular pressure (IOP) for severe glaucoma may require implantation of a glaucoma drainage device (GDD) to shunt aqueous humor (AH) from the anterior chamber of the eye and lower IOP to acceptable levels between 8 and 21 mm Hg. Nonvalved GDDs (NVGDDs) cannot maintain IOP in that acceptable range during the early postoperative period and require intra-operative modifications for IOP control during the first 30 days after surgery. Other GDDs have valves to overcome this issue, but are less successful with maintaining long-term IOP. Our research goal is to improve NVGDD postoperative performance. Little rigorous research has been done to systematically analyze flow/pressure characteristics in NVGDDs. We describe an experimental system developed to assess the pressure drop for physiologic flow rates through NVGDD-like microtubes of various lengths/diameters, some with annular inserts. Experimental pressure measurements for flow through hollow microtubes are within predictive theory's limits. For instance, a 50.4 μm inner diameter microtube yields an average experimental pressure of 33.7 mm Hg, while theory predicts 31.0–64.2 mm Hg. An annular example, with 358.8 μm outside and 330.7 μm inside diameters, yields an experimental pressure of 9.6 mm Hg, within theoretical predictions of 4.2–19.2 mm Hg. These results are repeatable and consistent over 25 days, which fits the 20–35 day period needed for scar tissue formation to achieve long-term IOP control. This work introduces a novel method for controlling IOP and demonstrates an experiment to examine this over 25 days. Future efforts will study insert size and degradable inserts.
publisherThe American Society of Mechanical Engineers (ASME)
titleValidation of an Experimental Setup to Reliably Simulate Flow Through Nonvalved Glaucoma Drainage Devices
typeJournal Paper
journal volume1
journal issue4
journal titleJournal of Engineering and Science in Medical Diagnostics and Therapy
identifier doi10.1115/1.4040498
journal fristpage41001
journal lastpage041001-8
treeJournal of Engineering and Science in Medical Diagnostics and Therapy:;2019:;volume( 001 ):;issue: 004
contenttypeFulltext