Elevated Blood Viscosity and Microrecirculation Resulting From Coronary Stent MalappositionSource: Journal of Biomechanical Engineering:;2018:;volume( 140 ):;issue: 005::page 51006Author:Poon, Eric K. W.
,
Thondapu, Vikas
,
Hayat, Umair
,
Barlis, Peter
,
Yap, Chooi Yin
,
Kuo, Po-Hung
,
Wang, Qisen
,
Ma, Jiawei
,
Zhu, Shuang J.
,
Moore, Stephen
,
Ooi, Andrew S. H.
DOI: 10.1115/1.4039306Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: One particular complexity of coronary artery is the natural tapering of the vessel with proximal segments having larger caliber and distal tapering as the vessel get smaller. The natural tapering of a coronary artery often leads to proximal incomplete stent apposition (ISA). ISA alters coronary hemodynamics and creates pathological path to develop complications such as in-stent restenosis, and more worryingly, stent thrombosis (ST). By employing state-of-the-art computer-aided design software, generic stent hoops were virtually deployed in an idealized tapered coronary artery with decreasing malapposition distance. Pulsatile blood flow simulations were carried out using computational fluid dynamics (CFD) on these computer-aided design models. CFD results reveal unprecedented details in both spatial and temporal development of microrecirculation environments throughout the cardiac cycle (CC). Arterial tapering also introduces secondary microrecirculation. These primary and secondary microrecirculations provoke significant fluctuations in arterial wall shear stress (WSS). There has been a direct correlation with changes in WSS and the development of atherosclerosis. Further, the presence of these microrecirculations influence strongly on the local levels of blood viscosity in the vicinity of the malapposed stent struts. The observation of secondary microrecirculations and changes in blood rheology is believed to complement the wall (-based) shear stress, perhaps providing additional physical explanations for tissue accumulation near ISA detected from high resolution optical coherence tomography (OCT).
|
Collections
Show full item record
| contributor author | Poon, Eric K. W. | |
| contributor author | Thondapu, Vikas | |
| contributor author | Hayat, Umair | |
| contributor author | Barlis, Peter | |
| contributor author | Yap, Chooi Yin | |
| contributor author | Kuo, Po-Hung | |
| contributor author | Wang, Qisen | |
| contributor author | Ma, Jiawei | |
| contributor author | Zhu, Shuang J. | |
| contributor author | Moore, Stephen | |
| contributor author | Ooi, Andrew S. H. | |
| date accessioned | 2019-02-28T11:11:14Z | |
| date available | 2019-02-28T11:11:14Z | |
| date copyright | 3/5/2018 12:00:00 AM | |
| date issued | 2018 | |
| identifier issn | 0148-0731 | |
| identifier other | bio_140_05_051006.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl1/handle/yetl/4253602 | |
| description abstract | One particular complexity of coronary artery is the natural tapering of the vessel with proximal segments having larger caliber and distal tapering as the vessel get smaller. The natural tapering of a coronary artery often leads to proximal incomplete stent apposition (ISA). ISA alters coronary hemodynamics and creates pathological path to develop complications such as in-stent restenosis, and more worryingly, stent thrombosis (ST). By employing state-of-the-art computer-aided design software, generic stent hoops were virtually deployed in an idealized tapered coronary artery with decreasing malapposition distance. Pulsatile blood flow simulations were carried out using computational fluid dynamics (CFD) on these computer-aided design models. CFD results reveal unprecedented details in both spatial and temporal development of microrecirculation environments throughout the cardiac cycle (CC). Arterial tapering also introduces secondary microrecirculation. These primary and secondary microrecirculations provoke significant fluctuations in arterial wall shear stress (WSS). There has been a direct correlation with changes in WSS and the development of atherosclerosis. Further, the presence of these microrecirculations influence strongly on the local levels of blood viscosity in the vicinity of the malapposed stent struts. The observation of secondary microrecirculations and changes in blood rheology is believed to complement the wall (-based) shear stress, perhaps providing additional physical explanations for tissue accumulation near ISA detected from high resolution optical coherence tomography (OCT). | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | Elevated Blood Viscosity and Microrecirculation Resulting From Coronary Stent Malapposition | |
| type | Journal Paper | |
| journal volume | 140 | |
| journal issue | 5 | |
| journal title | Journal of Biomechanical Engineering | |
| identifier doi | 10.1115/1.4039306 | |
| journal fristpage | 51006 | |
| journal lastpage | 051006-9 | |
| tree | Journal of Biomechanical Engineering:;2018:;volume( 140 ):;issue: 005 | |
| contenttype | Fulltext |