    Electromechanical Model-Based Design and Testing of Fiber Scanners for Endoscopy

    Source: Journal of Medical Devices:;2018:;volume( 012 ):;issue: 004::page 41003
    Author:
    Rajiv, Abhijith
    ,
    Zhou, Yaxuan
    ,
    Ridge, Jeremy
    ,
    Reinhall, Per G.
    ,
    Seibel, Eric J.
    DOI: 10.1115/1.4040271
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Forward-viewing catheters and scopes for diagnosing disease and guiding interventions in small ducts (less than 3 mm diameter) require wide-field high-quality imaging since scope tip bending is difficult and ineffective. A high-fidelity electromechanically coupled finite element (FE) model of a piezoelectric actuated resonant fiber scanner is presented, which enables improvement on the general design of fiber-optic scanner geometry to increase scan frequency and field of view (FOV). Using the proposed model, parametric sweeps on the specific design variables achieved by acid etching of glass fiber are analyzed to identify their effect on scanner performance and to choose improved designs. The resulting complex fiber scanner design requires development of unique microfabrication techniques. Comparison of three model simulations and their experimental testing show that our proposed coupled model has prediction error of ≤12% with respect to experimental data, while other uncoupled models have up to 39% error. The model and microfabrication techniques presented in this paper have significance for fiber scanning-based systems in that they demonstrate reliability for model-driven design and also flexibility for fiber scanner design of complex geometries, allowing for improvement on medical imaging performance.
    • Download: (1.851Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Electromechanical Model-Based Design and Testing of Fiber Scanners for Endoscopy

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4252472
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorRajiv, Abhijith
    contributor authorZhou, Yaxuan
    contributor authorRidge, Jeremy
    contributor authorReinhall, Per G.
    contributor authorSeibel, Eric J.
    date accessioned2019-02-28T11:04:54Z
    date available2019-02-28T11:04:54Z
    date copyright9/21/2018 12:00:00 AM
    date issued2018
    identifier issn1932-6181
    identifier othermed_012_04_041003.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252472
    description abstractForward-viewing catheters and scopes for diagnosing disease and guiding interventions in small ducts (less than 3 mm diameter) require wide-field high-quality imaging since scope tip bending is difficult and ineffective. A high-fidelity electromechanically coupled finite element (FE) model of a piezoelectric actuated resonant fiber scanner is presented, which enables improvement on the general design of fiber-optic scanner geometry to increase scan frequency and field of view (FOV). Using the proposed model, parametric sweeps on the specific design variables achieved by acid etching of glass fiber are analyzed to identify their effect on scanner performance and to choose improved designs. The resulting complex fiber scanner design requires development of unique microfabrication techniques. Comparison of three model simulations and their experimental testing show that our proposed coupled model has prediction error of ≤12% with respect to experimental data, while other uncoupled models have up to 39% error. The model and microfabrication techniques presented in this paper have significance for fiber scanning-based systems in that they demonstrate reliability for model-driven design and also flexibility for fiber scanner design of complex geometries, allowing for improvement on medical imaging performance.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleElectromechanical Model-Based Design and Testing of Fiber Scanners for Endoscopy
    typeJournal Paper
    journal volume12
    journal issue4
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4040271
    journal fristpage41003
    journal lastpage041003-8
    treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 004
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian