    Novel Design and Validation of a Micro Instrument in an Ear Grommet Insertion Device

    Source: Journal of Medical Devices:;2018:;volume( 012 ):;issue: 003::page 31004
    Author:
    Ng, Cailin
    ,
    Liang, Wenyu
    ,
    Gan, Chee Wee
    ,
    Lim, Hsueh Yee
    ,
    Tan, Kok Kiong
    DOI: 10.1115/1.4040185
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: An automated surgical device, the ventilator tube applicator (VTA), enables a grommet insertion surgery for patients with otitis media with effusion (OME) to be completed in a short time automatically and precisely, eliminating the use of general anesthesia (GA) typically required in such procedure. However, its current design limits the usefulness of the device as it is restricted by the properties of the tympanic membrane (TM), such as angle, thickness, and strength. Therefore, a novel design was conceptualized and the insertion control algorithm was improved to overcome the current challenges of the VTA. This innovative cover-cutter instrument design allows three-dimensional (3D) motion on an oblique surface using a single axis actuator. Experimental results on mock membranes showed great improvements in terms of robustness and success rate. The new design allowed the procedure to be performed on wider range of TM angles and hence increased the effectiveness of VTA. Grommet insertion force was reduced by an average of 66%, and the overall peak force reduced by an average of 14%. Finite element (FE) analysis on a cadaveric TM model further validated the usefulness of the cover-cutter instrument, and showed some interesting insights in the grommet insertion process.
    • Download: (3.275Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Novel Design and Validation of a Micro Instrument in an Ear Grommet Insertion Device

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4252467
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorNg, Cailin
    contributor authorLiang, Wenyu
    contributor authorGan, Chee Wee
    contributor authorLim, Hsueh Yee
    contributor authorTan, Kok Kiong
    date accessioned2019-02-28T11:04:53Z
    date available2019-02-28T11:04:53Z
    date copyright7/13/2018 12:00:00 AM
    date issued2018
    identifier issn1932-6181
    identifier othermed_012_03_031004.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252467
    description abstractAn automated surgical device, the ventilator tube applicator (VTA), enables a grommet insertion surgery for patients with otitis media with effusion (OME) to be completed in a short time automatically and precisely, eliminating the use of general anesthesia (GA) typically required in such procedure. However, its current design limits the usefulness of the device as it is restricted by the properties of the tympanic membrane (TM), such as angle, thickness, and strength. Therefore, a novel design was conceptualized and the insertion control algorithm was improved to overcome the current challenges of the VTA. This innovative cover-cutter instrument design allows three-dimensional (3D) motion on an oblique surface using a single axis actuator. Experimental results on mock membranes showed great improvements in terms of robustness and success rate. The new design allowed the procedure to be performed on wider range of TM angles and hence increased the effectiveness of VTA. Grommet insertion force was reduced by an average of 66%, and the overall peak force reduced by an average of 14%. Finite element (FE) analysis on a cadaveric TM model further validated the usefulness of the cover-cutter instrument, and showed some interesting insights in the grommet insertion process.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleNovel Design and Validation of a Micro Instrument in an Ear Grommet Insertion Device
    typeJournal Paper
    journal volume12
    journal issue3
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4040185
    journal fristpage31004
    journal lastpage031004-12
    treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 003
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian