Show simple item record

contributor authorPackett, Ryan D. M.
contributor authorBrown, Philip J.
contributor authorPopli, Gautam S. S.
contributor authorScott Gayzik, F.
date accessioned2017-11-25T07:19:14Z
date available2017-11-25T07:19:14Z
date copyright2017/6/4
date issued2017
identifier issn0148-0731
identifier otherbio_139_05_051007.pdf
identifier urihttp://138.201.223.254:8080/yetl1/handle/yetl/4235685
description abstractTissue cooling has been proven as a viable therapy for multiple conditions and injuries and has been applied to the brain to treat epilepsy and concussions, leading to improved long-term outcomes. To facilitate the study of temperature reduction as a function of various cooling methods, a thermal brain phantom was developed and analyzed. The phantom is composed of a potassium-neutralized, superabsorbent copolymer hydrogel. The phantom was tested in a series of cooling trials using a cooling block and 37 deg water representing nondirectional blood flow ranging up to 6 gph, a physiologically representative range based on the prototype volume. Results were compared against a validated finite difference (FD) model. Two sets of parameters were used in the FD model: one set to represent the phantom itself and a second set to represent brain parenchyma. The model was then used to calculate steady-state cooling at a depth of 5 mm for all flow rates, for both the phantom and a model of the brain. This effort was undertaken to (1) validate the FD model against the phantom results and (2) evaluate how similar the thermal response of the phantom is to that of a perfused brain. The FD phantom model showed good agreement with the empirical phantom results. Furthermore, the empirical phantom agreed with the predicted brain response within 3.5% at physiological flow, suggesting a biofidelic thermal response. The phantom will be used as a platform for future studies of thermally mediated therapies applied to the cerebral cortex.
publisherThe American Society of Mechanical Engineers (ASME)
titleDevelopment and Validation of a Brain Phantom for Therapeutic Cooling Devices
typeJournal Paper
journal volume139
journal issue5
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.4036215
journal fristpage51007
journal lastpage051007-7
treeJournal of Biomechanical Engineering:;2017:;volume( 139 ):;issue: 005
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record