    Computer-Controlled Eucapnic Voluntary Hyperpnea Challenge Method

    Source: Journal of Medical Devices:;2016:;volume( 010 ):;issue: 004::page 44507
    Author:
    Al-Jumaily, Ahmed M.
    ,
    Wang, Lulu
    DOI: 10.1115/1.4034146
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Eucapnic voluntary hyperpnea (EVH) challenge is a well-established and sensitive method of determining the degree of bronchoconstriction at hospitals and clinics. This paper presents the development of a computer-controlled system for assessing exercise-induced bronchoconstriction (EIB) in humans including a computer graphic user interface (GUI) to control a low-pressure demand valve for better efficiency. GUI is designed to monitor the severity of acute lung airway narrowing using a matlab software and to present the measurement data into a simple user-friendly program consisting of patient information, EVH test analysis, and detection of exercise-induced asthma (EIA) and EIB. The proposed system is assessed using human subjects. Typical outputs from this system showed that for a female participant, a 20.25% and a 15.61% decrease from baseline in her forced expiratory volume in 1 s (FEVl) after 10 and 15 min of the challenge commencement, respectively. Her actual expiratory flow rate (45.833 L/m) and actual total volume of gas respired (275 L) were smaller than the target values. This system widens the usage of EVH challenge in medical areas, and the GUI may serve as a new clinical computer-aided diagnostic tool to help healthcare professionals noninvasively monitor the severity of asthma, EIA, and EIB.
    • Download: (1.237Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Computer-Controlled Eucapnic Voluntary Hyperpnea Challenge Method

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4235181
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorAl-Jumaily, Ahmed M.
    contributor authorWang, Lulu
    date accessioned2017-11-25T07:18:27Z
    date available2017-11-25T07:18:27Z
    date copyright2016/09/12
    date issued2016
    identifier issn1932-6181
    identifier othermed_010_04_044507.pdf
    identifier urihttp://138.201.223.254:8080/yetl1/handle/yetl/4235181
    description abstractEucapnic voluntary hyperpnea (EVH) challenge is a well-established and sensitive method of determining the degree of bronchoconstriction at hospitals and clinics. This paper presents the development of a computer-controlled system for assessing exercise-induced bronchoconstriction (EIB) in humans including a computer graphic user interface (GUI) to control a low-pressure demand valve for better efficiency. GUI is designed to monitor the severity of acute lung airway narrowing using a matlab software and to present the measurement data into a simple user-friendly program consisting of patient information, EVH test analysis, and detection of exercise-induced asthma (EIA) and EIB. The proposed system is assessed using human subjects. Typical outputs from this system showed that for a female participant, a 20.25% and a 15.61% decrease from baseline in her forced expiratory volume in 1 s (FEVl) after 10 and 15 min of the challenge commencement, respectively. Her actual expiratory flow rate (45.833 L/m) and actual total volume of gas respired (275 L) were smaller than the target values. This system widens the usage of EVH challenge in medical areas, and the GUI may serve as a new clinical computer-aided diagnostic tool to help healthcare professionals noninvasively monitor the severity of asthma, EIA, and EIB.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleComputer-Controlled Eucapnic Voluntary Hyperpnea Challenge Method
    typeJournal Paper
    journal volume10
    journal issue4
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4034146
    journal fristpage44507
    journal lastpage044507-5
    treeJournal of Medical Devices:;2016:;volume( 010 ):;issue: 004
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian