Minimizing Human Tracking Error for Robotic Rehabilitation DeviceSource: Journal of Medical Devices:;2015:;volume( 009 ):;issue: 004::page 41003Author:Homich, Andrew J.
,
Doerzbacher, Megan A.
,
Tschantz, Eric L.
,
Piazza, Stephen J.
,
Hills, Everett C.
,
Moore, Jason Z.
DOI: 10.1115/1.4030275Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: This paper explores the design of a novel robotic device for gait training and rehabilitation, a method to estimate a human's orientation within the rehabilitation device, as well as an optimal state space controller to actuate the rehabilitation device. The robotic parallel bars (RPBs) were designed to address the shortcomings of currently available assistive devices. The RPB device moves in response to a human occupant to maintain a constant distance and orientation to the human. To minimize the error in tracking the human, a complementary filter was optimized to estimate the human's orientation within the device using a magnetometer and gyroscope. Experimental measurements of complementary filter performance on a test platform show that the filter estimates orientation with an average error of 0.62 deg over a range of angular velocities from 22.5 deg/s to 180 deg/s. The RPB device response was simulated, and an optimal state space controller was implemented using a linear quadratic regulator (LQR). The controller has an average position error of 14.1 cm and an average orientation error of 14.3 deg when tracking a human, while the simulation predicted an average error of 10.5 cm and 5.6 deg. The achieved level of accuracy in following a human user is sufficiently sensitive for the RPB device to conduct more advanced, realistic gait training and rehabilitation techniques for mobility impaired patients able to safely support their body weight with their legs and arms.
|
Collections
Show full item record
contributor author | Homich, Andrew J. | |
contributor author | Doerzbacher, Megan A. | |
contributor author | Tschantz, Eric L. | |
contributor author | Piazza, Stephen J. | |
contributor author | Hills, Everett C. | |
contributor author | Moore, Jason Z. | |
date accessioned | 2017-05-09T01:21:55Z | |
date available | 2017-05-09T01:21:55Z | |
date issued | 2015 | |
identifier issn | 1932-6181 | |
identifier other | med_009_04_041003.pdf | |
identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/159180 | |
description abstract | This paper explores the design of a novel robotic device for gait training and rehabilitation, a method to estimate a human's orientation within the rehabilitation device, as well as an optimal state space controller to actuate the rehabilitation device. The robotic parallel bars (RPBs) were designed to address the shortcomings of currently available assistive devices. The RPB device moves in response to a human occupant to maintain a constant distance and orientation to the human. To minimize the error in tracking the human, a complementary filter was optimized to estimate the human's orientation within the device using a magnetometer and gyroscope. Experimental measurements of complementary filter performance on a test platform show that the filter estimates orientation with an average error of 0.62 deg over a range of angular velocities from 22.5 deg/s to 180 deg/s. The RPB device response was simulated, and an optimal state space controller was implemented using a linear quadratic regulator (LQR). The controller has an average position error of 14.1 cm and an average orientation error of 14.3 deg when tracking a human, while the simulation predicted an average error of 10.5 cm and 5.6 deg. The achieved level of accuracy in following a human user is sufficiently sensitive for the RPB device to conduct more advanced, realistic gait training and rehabilitation techniques for mobility impaired patients able to safely support their body weight with their legs and arms. | |
publisher | The American Society of Mechanical Engineers (ASME) | |
title | Minimizing Human Tracking Error for Robotic Rehabilitation Device | |
type | Journal Paper | |
journal volume | 9 | |
journal issue | 4 | |
journal title | Journal of Medical Devices | |
identifier doi | 10.1115/1.4030275 | |
journal fristpage | 41003 | |
journal lastpage | 41003 | |
identifier eissn | 1932-619X | |
tree | Journal of Medical Devices:;2015:;volume( 009 ):;issue: 004 | |
contenttype | Fulltext |