Bioinspired Spring Loaded Biopsy Harvester—Experimental Prototype Design and Feasibility TestsSource: Journal of Medical Devices:;2014:;volume( 008 ):;issue: 001::page 15002DOI: 10.1115/1.4026449Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: Current minimally invasive laparoscopic tissue–harvesting techniques for pathological purposes involve taking multiple imprecise and inaccurate biopsies, usually using a laparoscopic forceps or other assistive devices. Potential hazards, e.g., cancer spread when dealing with tumorous tissue, call for a more reliable alternative in the form of a single laparoscopic instrument capable of repeatedly taking a precise biopsy at a desired location. Therefore, the aim of this project was to design a disposable laparoscopic instrument tip, incorporating a centrally positioned glass fiber for tissue diagnostics; a cutting device for fast, accurate, and reliable biopsy of a precisely defined volume; and a container suitable for sample storage. Inspired by the sea urchin's chewing organ, Aristotle's lantern, and its capability of rapid and simultaneous tissue incision and enclosure by axial translation, we designed a crownshaped collapsible cutter operating on a similar basis. Based on a series of in vitro experiments indicating that tissue deformation decreases with increasing penetration speed leading to a more precise biopsy, we decided on the cutter's forward propulsion via a spring. Apart from the embedded springloaded cutter, the biopsy harvester comprises a smart mechanism for cutter preloading, locking, and actuation, as well as a sample container. A realsized biopsy harvester prototype was developed and tested in a universal tensile testing machine at TU Delft. In terms of mechanical functionality, the preloading, locking, and actuation mechanism as well as the cutter's rapid incising and collapsing capabilities proved to work successfully in vitro. Further division of the tip into a permanent and a disposable segment will enable taking of multiple biopsies, mutually separated in individual containers. We believe the envisioned laparoscopic optomechanical biopsy device will be a solution ameliorating timedemanding, inaccurate, and potentially unsafe laparoscopic biopsy procedures.
|
Collections
Show full item record
| contributor author | Jelأnek, Filip | |
| contributor author | Smit, Gerwin | |
| contributor author | Breedveld, Paul | |
| date accessioned | 2017-05-09T01:11:07Z | |
| date available | 2017-05-09T01:11:07Z | |
| date issued | 2014 | |
| identifier issn | 1932-6181 | |
| identifier other | med_008_01_015002.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/155818 | |
| description abstract | Current minimally invasive laparoscopic tissue–harvesting techniques for pathological purposes involve taking multiple imprecise and inaccurate biopsies, usually using a laparoscopic forceps or other assistive devices. Potential hazards, e.g., cancer spread when dealing with tumorous tissue, call for a more reliable alternative in the form of a single laparoscopic instrument capable of repeatedly taking a precise biopsy at a desired location. Therefore, the aim of this project was to design a disposable laparoscopic instrument tip, incorporating a centrally positioned glass fiber for tissue diagnostics; a cutting device for fast, accurate, and reliable biopsy of a precisely defined volume; and a container suitable for sample storage. Inspired by the sea urchin's chewing organ, Aristotle's lantern, and its capability of rapid and simultaneous tissue incision and enclosure by axial translation, we designed a crownshaped collapsible cutter operating on a similar basis. Based on a series of in vitro experiments indicating that tissue deformation decreases with increasing penetration speed leading to a more precise biopsy, we decided on the cutter's forward propulsion via a spring. Apart from the embedded springloaded cutter, the biopsy harvester comprises a smart mechanism for cutter preloading, locking, and actuation, as well as a sample container. A realsized biopsy harvester prototype was developed and tested in a universal tensile testing machine at TU Delft. In terms of mechanical functionality, the preloading, locking, and actuation mechanism as well as the cutter's rapid incising and collapsing capabilities proved to work successfully in vitro. Further division of the tip into a permanent and a disposable segment will enable taking of multiple biopsies, mutually separated in individual containers. We believe the envisioned laparoscopic optomechanical biopsy device will be a solution ameliorating timedemanding, inaccurate, and potentially unsafe laparoscopic biopsy procedures. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | Bioinspired Spring Loaded Biopsy Harvester—Experimental Prototype Design and Feasibility Tests | |
| type | Journal Paper | |
| journal volume | 8 | |
| journal issue | 1 | |
| journal title | Journal of Medical Devices | |
| identifier doi | 10.1115/1.4026449 | |
| journal fristpage | 15002 | |
| journal lastpage | 15002 | |
| identifier eissn | 1932-619X | |
| tree | Journal of Medical Devices:;2014:;volume( 008 ):;issue: 001 | |
| contenttype | Fulltext |