    A Comparison of the Temperature Rise Generated in Bone by the Use of a Standard Oscillating Saw Blade and the “Precisionâ€‌ Saw Blade

    Source: Journal of Medical Devices:;2013:;volume( 007 ):;issue: 002::page 21006
    Author:
    Lustig, Sebastien
    ,
    Scholes, Corey J.
    ,
    Oussedik, Sam
    ,
    Tam, Sam
    ,
    Dabirrahmani, Danأ¨
    ,
    Appleyard, Richard
    ,
    Parker, David A.
    DOI: 10.1115/1.4024159
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Introduction. Osteonecrosis may be triggered by bone temperatures above 47 آ°C during routine orthopaedic bone cuts using powerdriven saws with potentially negative impacts on bone healing. A new oscillatingtip saw blade design (Precisionآ®; Stryker, Kalamazoo, USA) has been recently developed with a saw blade design that may influence the amount of heat generated. We have, therefore, sought to compare the bone temperature achieved using this new blade design with a standard oscillating saw during a standardized cutting task. Method. Six human cadaveric femora were obtained. Each femur was clamped and a distal femoral cutting jig was applied. An initial cut was performed to visualize the distal metaphyseal bone. The cutting block was then moved 2 mm proximal and a further cut performed, measuring the temperature of the bone with an infrared camera. This was repeated, moving the block 2 mm proximal with each cut, alternating between a standard oscillating saw blade (12 cuts) and the Precisionآ® saw blade (12 cuts). The bone density at the level of each slice was established from a CT scan of each specimen which had been performed prior to the experiment. Results. The two blades did not differ with respect to the integrated mean temperature calculated for each cut (p = 0.89). The average peak temperatures were not significantly different between blades (p = 0.14). There was no significant difference between blades for peak heating rate (p = 0.7), although the area of bone heated above the 47 deg osteonecrotic threshold was significantly (p = 0.04) less for the standard saw blade. Conclusions. The Precisionآ® blade may have advantages over standard oscillating blade, but reduced heat generation was not observed in this study. Indeed, the Precisionآ® blade generated heat that exceeded the bony osteonecrosis threshold in a greater proportion of bone than the standard blade, questioning its use for osteotomy or uncemented knee arthroplasty. Further work should examine modifications to the blade design to better optimize the requirements of speed, accuracy and heat generation.
    • Download: (424.4Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      A Comparison of the Temperature Rise Generated in Bone by the Use of a Standard Oscillating Saw Blade and the “Precisionâ€‌ Saw Blade

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/152732
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorLustig, Sebastien
    contributor authorScholes, Corey J.
    contributor authorOussedik, Sam
    contributor authorTam, Sam
    contributor authorDabirrahmani, Danأ¨
    contributor authorAppleyard, Richard
    contributor authorParker, David A.
    date accessioned2017-05-09T01:01:30Z
    date available2017-05-09T01:01:30Z
    date issued2013
    identifier issn1932-6181
    identifier othermed_7_2_021006.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/152732
    description abstractIntroduction. Osteonecrosis may be triggered by bone temperatures above 47 آ°C during routine orthopaedic bone cuts using powerdriven saws with potentially negative impacts on bone healing. A new oscillatingtip saw blade design (Precisionآ®; Stryker, Kalamazoo, USA) has been recently developed with a saw blade design that may influence the amount of heat generated. We have, therefore, sought to compare the bone temperature achieved using this new blade design with a standard oscillating saw during a standardized cutting task. Method. Six human cadaveric femora were obtained. Each femur was clamped and a distal femoral cutting jig was applied. An initial cut was performed to visualize the distal metaphyseal bone. The cutting block was then moved 2 mm proximal and a further cut performed, measuring the temperature of the bone with an infrared camera. This was repeated, moving the block 2 mm proximal with each cut, alternating between a standard oscillating saw blade (12 cuts) and the Precisionآ® saw blade (12 cuts). The bone density at the level of each slice was established from a CT scan of each specimen which had been performed prior to the experiment. Results. The two blades did not differ with respect to the integrated mean temperature calculated for each cut (p = 0.89). The average peak temperatures were not significantly different between blades (p = 0.14). There was no significant difference between blades for peak heating rate (p = 0.7), although the area of bone heated above the 47 deg osteonecrotic threshold was significantly (p = 0.04) less for the standard saw blade. Conclusions. The Precisionآ® blade may have advantages over standard oscillating blade, but reduced heat generation was not observed in this study. Indeed, the Precisionآ® blade generated heat that exceeded the bony osteonecrosis threshold in a greater proportion of bone than the standard blade, questioning its use for osteotomy or uncemented knee arthroplasty. Further work should examine modifications to the blade design to better optimize the requirements of speed, accuracy and heat generation.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleA Comparison of the Temperature Rise Generated in Bone by the Use of a Standard Oscillating Saw Blade and the “Precisionâ€‌ Saw Blade
    typeJournal Paper
    journal volume7
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4024159
    journal fristpage21006
    journal lastpage21006
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2013:;volume( 007 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian