    Computational Fluid Dynamics Evaluation of Equivalency in Hemodynamic Alterations Between Driver, Integrity, and Similar Stents Implanted Into an Idealized Coronary Artery

    Source: Journal of Medical Devices:;2013:;volume( 007 ):;issue: 001::page 11004
    Author:
    Gundert, Timothy J.
    ,
    Dholakia, Ronak J.
    ,
    McMahon, Dennis
    ,
    LaDisa, , Jr. ,John F.
    DOI: 10.1115/1.4023413
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: We tested the hypothesis that a slight modification in fabrication from the Driver to the Integrity stent (Medtronic) results in nearly equivalent distributions of wall shear stress (WSS) and mean exposure time (MET), reflective of flow stagnation, and that these differences are considerably less than the MultiLink Vision (Abbott Vascular) or BX Velocity (Cordis) bare metal stents when evaluated by computational fluid dynamics (CFD). Arteries were modeled as idealized straight rigid vessels without lesions. Two vessel diameters (2.25 and 3.0 mm) were studied for each stent and 2.75 mm diameter Integrity stents were also modeled to quantify the impact from bestand worstcase orientations of the stent struts relative to the primary blood flow direction. All stents were 18 mm in length and overdeployed by 10%. The results indicated that, regardless of diameter, the BX Velocity stents had the greatest percentage of the vessel exposed to adverse WSS followed by the Vision, Integrity, and Driver stents. In general, when strut thickness and stent:lumen ratio are similar, the orientation of struts is a determining factor for deleterious flow patterns. For a given stent, the number of struts was a larger determinant of adverse WSS and MET than strut orientation, suggesting that favorable blood flow patterns can be achieved by limiting struts to those providing adequate scaffolding. In conclusion, the Driver and Integrity stents both limit their number of linkages to those which provide adequate scaffolding while also maintaining similar strut thickness and stent:lumen ratios. The Integrity stent also imparts a slight helical velocity component. The modest difference in the fabrication approach between the Driver and Integrity stents is, therefore, not hemodynamically substantial in this idealized analysis, particularly relative to potentially adverse flow conditions introduced by the other stents modeled. This data was used in conjunction with associated regulatory filings and submitted to the FDA as part of the documents facilitating the recent approval for sale of the Resolute Integrity stent in the United States.
    • Download: (5.972Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Computational Fluid Dynamics Evaluation of Equivalency in Hemodynamic Alterations Between Driver, Integrity, and Similar Stents Implanted Into an Idealized Coronary Artery

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/152663
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorGundert, Timothy J.
    contributor authorDholakia, Ronak J.
    contributor authorMcMahon, Dennis
    contributor authorLaDisa, , Jr. ,John F.
    date accessioned2017-05-09T01:01:22Z
    date available2017-05-09T01:01:22Z
    date issued2013
    identifier issn1932-6181
    identifier othermed_7_1_011004.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/152663
    description abstractWe tested the hypothesis that a slight modification in fabrication from the Driver to the Integrity stent (Medtronic) results in nearly equivalent distributions of wall shear stress (WSS) and mean exposure time (MET), reflective of flow stagnation, and that these differences are considerably less than the MultiLink Vision (Abbott Vascular) or BX Velocity (Cordis) bare metal stents when evaluated by computational fluid dynamics (CFD). Arteries were modeled as idealized straight rigid vessels without lesions. Two vessel diameters (2.25 and 3.0 mm) were studied for each stent and 2.75 mm diameter Integrity stents were also modeled to quantify the impact from bestand worstcase orientations of the stent struts relative to the primary blood flow direction. All stents were 18 mm in length and overdeployed by 10%. The results indicated that, regardless of diameter, the BX Velocity stents had the greatest percentage of the vessel exposed to adverse WSS followed by the Vision, Integrity, and Driver stents. In general, when strut thickness and stent:lumen ratio are similar, the orientation of struts is a determining factor for deleterious flow patterns. For a given stent, the number of struts was a larger determinant of adverse WSS and MET than strut orientation, suggesting that favorable blood flow patterns can be achieved by limiting struts to those providing adequate scaffolding. In conclusion, the Driver and Integrity stents both limit their number of linkages to those which provide adequate scaffolding while also maintaining similar strut thickness and stent:lumen ratios. The Integrity stent also imparts a slight helical velocity component. The modest difference in the fabrication approach between the Driver and Integrity stents is, therefore, not hemodynamically substantial in this idealized analysis, particularly relative to potentially adverse flow conditions introduced by the other stents modeled. This data was used in conjunction with associated regulatory filings and submitted to the FDA as part of the documents facilitating the recent approval for sale of the Resolute Integrity stent in the United States.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleComputational Fluid Dynamics Evaluation of Equivalency in Hemodynamic Alterations Between Driver, Integrity, and Similar Stents Implanted Into an Idealized Coronary Artery
    typeJournal Paper
    journal volume7
    journal issue1
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4023413
    journal fristpage11004
    journal lastpage11004
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2013:;volume( 007 ):;issue: 001
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian