    A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis Index

    Source: Journal of Biomechanical Engineering:;2012:;volume( 134 ):;issue: 008::page 81002
    Author:
    Katharine H. Fraser
    ,
    Tao Zhang
    ,
    M. Ertan Taskin
    ,
    Bartley P. Griffith
    ,
    Zhongjun J. Wu
    DOI: 10.1115/1.4007092
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Ventricular assist devices (VADs) have already helped many patients with heart failure but have the potential to assist more patients if current problems with blood damage (hemolysis, platelet activation, thrombosis and emboli, and destruction of the von Willebrand factor (vWf)) can be eliminated. A step towards this goal is better understanding of the relationships between shear stress, exposure time, and blood damage and, from there, the development of numerical models for the different types of blood damage to enable the design of improved VADs. In this study, computational fluid dynamics (CFD) was used to calculate the hemodynamics in three clinical VADs and two investigational VADs and the shear stress, residence time, and hemolysis were investigated. A new scalar transport model for hemolysis was developed. The results were compared with in vitro measurements of the pressure head in each VAD and the hemolysis index in two VADs. A comparative analysis of the blood damage related fluid dynamic parameters and hemolysis index was performed among the VADs. Compared to the centrifugal VADs, the axial VADs had: higher mean scalar shear stress (sss); a wider range of sss, with larger maxima and larger percentage volumes at both low and high sss; and longer residence times at very high sss. The hemolysis predictions were in agreement with the experiments and showed that the axial VADs had a higher hemolysis index. The increased hemolysis in axial VADs compared to centrifugal VADs is a direct result of their higher shear stresses and longer residence times. Since platelet activation and destruction of the vWf also require high shear stresses, the flow conditions inside axial VADs are likely to result in more of these types of blood damage compared with centrifugal VADs.
    keyword(s): Flow (Dynamics) , Stress , Shear (Mechanics) , Blood , Pressure , Scalars AND Ventricular assist devices ,
    • Download: (2.797Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis Index

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/148218
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorKatharine H. Fraser
    contributor authorTao Zhang
    contributor authorM. Ertan Taskin
    contributor authorBartley P. Griffith
    contributor authorZhongjun J. Wu
    date accessioned2017-05-09T00:48:24Z
    date available2017-05-09T00:48:24Z
    date copyrightAugust, 2012
    date issued2012
    identifier issn0148-0731
    identifier otherJBENDY-29000#081002_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/148218
    description abstractVentricular assist devices (VADs) have already helped many patients with heart failure but have the potential to assist more patients if current problems with blood damage (hemolysis, platelet activation, thrombosis and emboli, and destruction of the von Willebrand factor (vWf)) can be eliminated. A step towards this goal is better understanding of the relationships between shear stress, exposure time, and blood damage and, from there, the development of numerical models for the different types of blood damage to enable the design of improved VADs. In this study, computational fluid dynamics (CFD) was used to calculate the hemodynamics in three clinical VADs and two investigational VADs and the shear stress, residence time, and hemolysis were investigated. A new scalar transport model for hemolysis was developed. The results were compared with in vitro measurements of the pressure head in each VAD and the hemolysis index in two VADs. A comparative analysis of the blood damage related fluid dynamic parameters and hemolysis index was performed among the VADs. Compared to the centrifugal VADs, the axial VADs had: higher mean scalar shear stress (sss); a wider range of sss, with larger maxima and larger percentage volumes at both low and high sss; and longer residence times at very high sss. The hemolysis predictions were in agreement with the experiments and showed that the axial VADs had a higher hemolysis index. The increased hemolysis in axial VADs compared to centrifugal VADs is a direct result of their higher shear stresses and longer residence times. Since platelet activation and destruction of the vWf also require high shear stresses, the flow conditions inside axial VADs are likely to result in more of these types of blood damage compared with centrifugal VADs.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleA Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis Index
    typeJournal Paper
    journal volume134
    journal issue8
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4007092
    journal fristpage81002
    identifier eissn1528-8951
    keywordsFlow (Dynamics)
    keywordsStress
    keywordsShear (Mechanics)
    keywordsBlood
    keywordsPressure
    keywordsScalars AND Ventricular assist devices
    treeJournal of Biomechanical Engineering:;2012:;volume( 134 ):;issue: 008
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian