contributor authorA. Changoor
contributor authorM. Garon
contributor authorM. B. Hurtig
contributor authorM. D. Buschmann
contributor authorE. Quenneville
contributor authorJ. P. Coutu
date accessioned2017-05-09T00:42:28Z
date available2017-05-09T00:42:28Z
date copyrightJune, 2011
date issued2011
identifier issn0148-0731
identifier otherJBENDY-27209#061005_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/145429
description abstractModels of post-traumatic osteoarthritis where early degenerative changes can be monitored are valuable for assessing potential therapeutic strategies. Current methods for evaluating cartilage mechanical properties may not capture the low-grade cartilage changes expected at these earlier time points following injury. In this study, an explant model of cartilage injury was used to determine whether streaming potential measurements by manual indentation could detect cartilage changes immediately following mechanical impact and to compare their sensitivity to biomechanical tests. Impacts were delivered ex vivo, at one of three stress levels, to specific positions on isolated adult equine trochlea. Cartilage properties were assessed by streaming potential measurements, made pre- and post-impact using a commercially available arthroscopic device, and by stress relaxation tests in unconfined compression geometry of isolated cartilage disks, providing the streaming potential integral (SPI), fibril modulus (Ef), matrix modulus (Em), and permeability (k). Histological sections were stained with Safranin-O and adjacent unstained sections examined in polarized light microscopy. Impacts were low, 17.3 ± 2.7 MPa (n = 15), medium, 27.8 ± 8.5 MPa (n = 13), or high, 48.7 ± 12.1 MPa (n = 16), and delivered using a custom-built spring-loaded device with a rise time of approximately 1 ms. SPI was significantly reduced after medium (p = 0.006) and high (p<0.001) impacts. Ef, representing collagen network stiffness, was significantly reduced in high impact samples only (p < 0.001 lateral trochlea, p = 0.042 medial trochlea), where permeability also increased (p = 0.003 lateral trochlea, p = 0.007 medial trochlea). Significant (p < 0.05, n = 68) moderate to strong correlations between SPI and Ef (r = 0.857), Em (r = 0.493), log(k) (r = −0.484), and cartilage thickness (r = −0.804) were detected. Effect sizes were higher for SPI than Ef, Em, and k, indicating greater sensitivity of electromechanical measurements to impact injury compared to purely biomechanical parameters. Histological changes due to impact were limited to the presence of superficial zone damage which increased with impact stress. Non-destructive streaming potential measurements were more sensitive to impact-related articular cartilage changes than biomechanical assessment of isolated samples using stress relaxation tests in unconfined compression geometry. Correlations between electromechanical and biomechanical methods further support the relationship between non-destructive electromechanical measurements and intrinsic cartilage properties.
publisherThe American Society of Mechanical Engineers (ASME)
titleStreaming Potential-Based Arthroscopic Device is Sensitive to Cartilage Changes Immediately Post-Impact in an Equine Cartilage Injury Model
typeJournal Paper
journal volume133
journal issue6
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.4004230
journal fristpage61005
identifier eissn1528-8951
keywordsMeasurement
keywordsBiomechanics
keywordsTesting
keywordsCompression
keywordsCartilage
keywordsStress
keywordsWounds
keywordsArthroscopy AND Disks
treeJournal of Biomechanical Engineering:;2011:;volume( 133 ):;issue: 006
contenttypeFulltext