    Disinfection of Male Luer Style Connectors for Prevention of Catheter Related Bloodstream Infections Using an Isopropyl Alcohol Dispensing Cap

    Source: Journal of Medical Devices:;2010:;volume( 004 ):;issue: 002::page 27509
    Author:
    James P. Kennedy
    ,
    Richard A. Lasher
    ,
    Donald Solomon
    ,
    Robert W. Hitchcock
    DOI: 10.1115/1.3442645
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Bacterial colonization of needleless injection sites (NISs) frequently results in catheter related bloodstream infections (CRBSIs). Hospitals have instituted protocols aimed at disinfecting NIS prior to access. Furthermore, several manufactures have developed devices that facilitate disinfection of NIS. Despite these steps, the incidence of CRBSI is still alarmingly high. Currently, there is no protocol or device intended to disinfect male luer connectors such as those found on IV tubing that are commonly coupled and decoupled from the NISs. Since these IV tubing connectors directly contact the NIS (which have been repeatedly shown to have varying levels of bacterial colonization), it is highly likely that they, too, will have varying levels of contamination. In order for disinfection of the NIS to be effective, the IV tubing connector must also be disinfected. Our design goal was to develop a device that could be used to disinfect a male luer style connector without allowing antiseptic into the inner lumen of the male luer. We designed a three component system that utilizes a silicone sealing cone to seal the male luer, a reservoir foam that holds 70% isopropyl alcohol (IPA), and a reaction force foam that increases the seal pressure of the sealing cone while the reservoir foam is compressed delivering the IPA to the outside surface of the male luer post. Sealing cone geometry was optimized using a custom built seal pressure test apparatus. Reservoir and reaction force foam functional parameters were assessed using an Instron test apparatus. A two phase compression stroke was designed into the device to allow for sealing and dispensing of IPA. An IPA transfer test was used to assess the transfer of disinfectant from the reservoir foam to a liquid filled male luer connector (modeling an IV tubing connector). No disinfectant was found to be transferred from the device to the inner lumen of the IV tubing connector model (n=30). To test the efficacy of the device on reducing bacterial count on the male luer, a disinfection study was performed using the optimized device. Male luers were immersed in bacterial suspensions of S. aureus, S. epidermis, P. aerginosa, and E. coli. A 4 log reduction compared with a positive control was found in each sample treated with our disinfection cap (n=120). In conclusion, we developed a device that effectively delivers an antiseptic to a male luer style connector without leaking any antiseptic to the inner lumen of the luer post
    • Download: (32.59Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Disinfection of Male Luer Style Connectors for Prevention of Catheter Related Bloodstream Infections Using an Isopropyl Alcohol Dispensing Cap

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/144408
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorJames P. Kennedy
    contributor authorRichard A. Lasher
    contributor authorDonald Solomon
    contributor authorRobert W. Hitchcock
    date accessioned2017-05-09T00:40:00Z
    date available2017-05-09T00:40:00Z
    date copyrightJune, 2010
    date issued2010
    identifier issn1932-6181
    identifier otherJMDOA4-28010#027509_2.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/144408
    description abstractBacterial colonization of needleless injection sites (NISs) frequently results in catheter related bloodstream infections (CRBSIs). Hospitals have instituted protocols aimed at disinfecting NIS prior to access. Furthermore, several manufactures have developed devices that facilitate disinfection of NIS. Despite these steps, the incidence of CRBSI is still alarmingly high. Currently, there is no protocol or device intended to disinfect male luer connectors such as those found on IV tubing that are commonly coupled and decoupled from the NISs. Since these IV tubing connectors directly contact the NIS (which have been repeatedly shown to have varying levels of bacterial colonization), it is highly likely that they, too, will have varying levels of contamination. In order for disinfection of the NIS to be effective, the IV tubing connector must also be disinfected. Our design goal was to develop a device that could be used to disinfect a male luer style connector without allowing antiseptic into the inner lumen of the male luer. We designed a three component system that utilizes a silicone sealing cone to seal the male luer, a reservoir foam that holds 70% isopropyl alcohol (IPA), and a reaction force foam that increases the seal pressure of the sealing cone while the reservoir foam is compressed delivering the IPA to the outside surface of the male luer post. Sealing cone geometry was optimized using a custom built seal pressure test apparatus. Reservoir and reaction force foam functional parameters were assessed using an Instron test apparatus. A two phase compression stroke was designed into the device to allow for sealing and dispensing of IPA. An IPA transfer test was used to assess the transfer of disinfectant from the reservoir foam to a liquid filled male luer connector (modeling an IV tubing connector). No disinfectant was found to be transferred from the device to the inner lumen of the IV tubing connector model (n=30). To test the efficacy of the device on reducing bacterial count on the male luer, a disinfection study was performed using the optimized device. Male luers were immersed in bacterial suspensions of S. aureus, S. epidermis, P. aerginosa, and E. coli. A 4 log reduction compared with a positive control was found in each sample treated with our disinfection cap (n=120). In conclusion, we developed a device that effectively delivers an antiseptic to a male luer style connector without leaking any antiseptic to the inner lumen of the luer post
    publisherThe American Society of Mechanical Engineers (ASME)
    titleDisinfection of Male Luer Style Connectors for Prevention of Catheter Related Bloodstream Infections Using an Isopropyl Alcohol Dispensing Cap
    typeJournal Paper
    journal volume4
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.3442645
    journal fristpage27509
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2010:;volume( 004 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian