    Repair of Periprosthetic Pelvis Defects With Porous Metal Implants: A Finite Element Study

    Source: Journal of Biomechanical Engineering:;2010:;volume( 132 ):;issue: 002::page 21006
    Author:
    Eik Siggelkow
    ,
    Roy D. Crowninshield
    ,
    Dale A. Degroff
    ,
    Danny L. Levine
    ,
    Mehul A. Dharia
    ,
    Douglas H. Wentz
    DOI: 10.1115/1.4000853
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Periacetabular osteolysis is a potentially difficult surgical challenge, which can often drive the choice of reconstruction methods used in revision hip replacement. For smaller defects, impaction of bone grafts may be sufficient, but larger defects can require filler materials that provide structural support in addition to filling a void. This study utilized finite element analysis (FEA) to examine the state of stress in periprosthetic pelvic bone when subjected to a stair-climbing load and in the presence of two simulated defects, to show the effect of implanting a defect repair implant fabricated from Trabecular Metal™ . Even a small medial bone defect showed a local stress elevation of 4× compared with that seen with an acetabular implant supported by intact periacetabular bone. Local bone stress was much greater (8× the baseline level) for a defect case in which the loss of bone superior to the acetabular implant permitted significant migration. FEA results showed that a repair of the small defect with a Trabecular Metal™ restrictor lowered periprosthetic bone stress to a level comparable to that in the case of a primary implant. For the larger defect case, the use of a Trabecular Metal™ augment provides structural stabilization and helps to restore the THR head center. However, stress in the adjacent periprosthetic bone is lower than that observed in the defect-free acetabulum. In the augment case, the load path between the femoral head and the pelvis now passes through the augment as the superior rim of the acetabulum has been replaced. Contact-induced stress in the augment is similar in magnitude to that seen in the superior rim of the baseline case, although the stress pattern in the augment is noticeably different from that in intact bone.
    keyword(s): Metals , Maintenance , Product quality , Stress , Bone , Finite element analysis , Surgery AND Fillers (Materials) ,
    • Download: (955.9Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Repair of Periprosthetic Pelvis Defects With Porous Metal Implants: A Finite Element Study

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/142670
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorEik Siggelkow
    contributor authorRoy D. Crowninshield
    contributor authorDale A. Degroff
    contributor authorDanny L. Levine
    contributor authorMehul A. Dharia
    contributor authorDouglas H. Wentz
    date accessioned2017-05-09T00:36:42Z
    date available2017-05-09T00:36:42Z
    date copyrightFebruary, 2010
    date issued2010
    identifier issn0148-0731
    identifier otherJBENDY-27104#021006_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/142670
    description abstractPeriacetabular osteolysis is a potentially difficult surgical challenge, which can often drive the choice of reconstruction methods used in revision hip replacement. For smaller defects, impaction of bone grafts may be sufficient, but larger defects can require filler materials that provide structural support in addition to filling a void. This study utilized finite element analysis (FEA) to examine the state of stress in periprosthetic pelvic bone when subjected to a stair-climbing load and in the presence of two simulated defects, to show the effect of implanting a defect repair implant fabricated from Trabecular Metal™ . Even a small medial bone defect showed a local stress elevation of 4× compared with that seen with an acetabular implant supported by intact periacetabular bone. Local bone stress was much greater (8× the baseline level) for a defect case in which the loss of bone superior to the acetabular implant permitted significant migration. FEA results showed that a repair of the small defect with a Trabecular Metal™ restrictor lowered periprosthetic bone stress to a level comparable to that in the case of a primary implant. For the larger defect case, the use of a Trabecular Metal™ augment provides structural stabilization and helps to restore the THR head center. However, stress in the adjacent periprosthetic bone is lower than that observed in the defect-free acetabulum. In the augment case, the load path between the femoral head and the pelvis now passes through the augment as the superior rim of the acetabulum has been replaced. Contact-induced stress in the augment is similar in magnitude to that seen in the superior rim of the baseline case, although the stress pattern in the augment is noticeably different from that in intact bone.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleRepair of Periprosthetic Pelvis Defects With Porous Metal Implants: A Finite Element Study
    typeJournal Paper
    journal volume132
    journal issue2
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4000853
    journal fristpage21006
    identifier eissn1528-8951
    keywordsMetals
    keywordsMaintenance
    keywordsProduct quality
    keywordsStress
    keywordsBone
    keywordsFinite element analysis
    keywordsSurgery AND Fillers (Materials)
    treeJournal of Biomechanical Engineering:;2010:;volume( 132 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian