    Numerical Study of Shear-Induced Thrombus Formation Over Aterial Stent Struts

    Source: Journal of Medical Devices:;2009:;volume( 003 ):;issue: 002::page 27525
    Author:
    D. L. Bark
    ,
    D. N. Ku
    DOI: 10.1115/1.3147259
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Thrombosis remains an important problem in both bare-metal and drug eluting stents. Platelet accumulation appears to be highly shear dependent in uniform parallel plate and stenosis models. This study was performed to evaluate thrombus formation location and size with respect to time for a stent. A three-dimensional laminar flow field was modeled via computational fluid dynamics through a stent-containing coronary sized vessel. Platelet deposition and accumulation was then simulated using a shear dependent accumulation function. The platelet deposition rate is given by the function: 3.2γ̇+67 platelets/mm2 /s, where γ̇ is the shear rate, which comes from a linear regression to data presented in Ku and Flannery 2007. A three-dimensional stent with a helical strut matrix design and a pitch of 21 mm was designed around a 3 mm diameter vessel. Square strut designs of 0.15mm×0.15mm and 0.30mm×0.30mm were considered, with each strut embedded halfway into the vessel. A 30° section of the stent was modeled because of stent symmetry. The inlet was set at a mean Reynolds number of 200 by specifying a pressure differential across the length of the vessel. Thrombus growth based on shear rate was set through the equation: dΦ/dt=(3.2γ̇+67)(Vplatelet/V)∑n=0MAn. V is volume, Φ is the volume fraction of thrombus, M is the number of surrounding faces that are either a wall face or are neighboring a computational cell denoted as occluded by thrombus, associated with area, A. Each term without a subscript pertains to the local computational cell prescribed as undergoing thrombus growth. Thrombus cell occlusion was assumed to occur when thrombus filled 80% of the cell's volume. Maximum shear rates occur near the edges along the inner blood surface side of the stent struts. Thrombus growth increases more in the axial direction of the stent relative to the radial direction for the larger strut size. Conversely, the thrombus growth is more uniform in all directions along the smaller struts. The difference emanates from the two distinctly localized high shear contours near the edges of the larger stent struts, while the high shear regions were less distinct for the smaller strut size. Quadrupling the cross sectional area of a strut increases the initial maximum shear rate along the strut by 50%, in addition to doubling the available surface area for platelet deposition. Therefore, increasing strut size has a threefold effect on platelet deposition rate, leading to faster vessel occlusion. This computational technique may be extended to approximate where thrombus may grow to completely occlude a blood vessel and the length of time occlusion would take. The CFD modeling technique may also be used to evaluate thrombus deposition on medical devices such as heart valves and ventricular assist devices.
    • Download: (39.09Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Numerical Study of Shear-Induced Thrombus Formation Over Aterial Stent Struts

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/141593
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorD. L. Bark
    contributor authorD. N. Ku
    date accessioned2017-05-09T00:34:44Z
    date available2017-05-09T00:34:44Z
    date copyrightJune, 2009
    date issued2009
    identifier issn1932-6181
    identifier otherJMDOA4-28002#027525_2.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/141593
    description abstractThrombosis remains an important problem in both bare-metal and drug eluting stents. Platelet accumulation appears to be highly shear dependent in uniform parallel plate and stenosis models. This study was performed to evaluate thrombus formation location and size with respect to time for a stent. A three-dimensional laminar flow field was modeled via computational fluid dynamics through a stent-containing coronary sized vessel. Platelet deposition and accumulation was then simulated using a shear dependent accumulation function. The platelet deposition rate is given by the function: 3.2γ̇+67 platelets/mm2 /s, where γ̇ is the shear rate, which comes from a linear regression to data presented in Ku and Flannery 2007. A three-dimensional stent with a helical strut matrix design and a pitch of 21 mm was designed around a 3 mm diameter vessel. Square strut designs of 0.15mm×0.15mm and 0.30mm×0.30mm were considered, with each strut embedded halfway into the vessel. A 30° section of the stent was modeled because of stent symmetry. The inlet was set at a mean Reynolds number of 200 by specifying a pressure differential across the length of the vessel. Thrombus growth based on shear rate was set through the equation: dΦ/dt=(3.2γ̇+67)(Vplatelet/V)∑n=0MAn. V is volume, Φ is the volume fraction of thrombus, M is the number of surrounding faces that are either a wall face or are neighboring a computational cell denoted as occluded by thrombus, associated with area, A. Each term without a subscript pertains to the local computational cell prescribed as undergoing thrombus growth. Thrombus cell occlusion was assumed to occur when thrombus filled 80% of the cell's volume. Maximum shear rates occur near the edges along the inner blood surface side of the stent struts. Thrombus growth increases more in the axial direction of the stent relative to the radial direction for the larger strut size. Conversely, the thrombus growth is more uniform in all directions along the smaller struts. The difference emanates from the two distinctly localized high shear contours near the edges of the larger stent struts, while the high shear regions were less distinct for the smaller strut size. Quadrupling the cross sectional area of a strut increases the initial maximum shear rate along the strut by 50%, in addition to doubling the available surface area for platelet deposition. Therefore, increasing strut size has a threefold effect on platelet deposition rate, leading to faster vessel occlusion. This computational technique may be extended to approximate where thrombus may grow to completely occlude a blood vessel and the length of time occlusion would take. The CFD modeling technique may also be used to evaluate thrombus deposition on medical devices such as heart valves and ventricular assist devices.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleNumerical Study of Shear-Induced Thrombus Formation Over Aterial Stent Struts
    typeJournal Paper
    journal volume3
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.3147259
    journal fristpage27525
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2009:;volume( 003 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian