    An Inexpensive Weight Bearing Indicator With Load Range Capability for Rehabilitation of Patients With Lower Extremity Injuries

    Source: Journal of Medical Devices:;2009:;volume( 003 ):;issue: 003::page 31001
    Author:
    Daniel F. Walczyk
    ,
    William T. Ziomek
    DOI: 10.1115/1.3212557
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: This paper investigates the mechanical behavior and design of a patent pending device called a load range weight bearing indicator (LWBI), which provides upper and lower range indication to patients with lower extremity injuries as part of a partial weight bearing rehabilitation. The LWBI consists of two opposing stacks (a.k.a. double stack) of snap domes—bistable mechanical elements that snap through only when a threshold weight is applied—sandwiched between a load transfer plate and base plate. The mechanical behavior of a LWBI has been characterized by testing single and double stacks of snap domes in a rigid aluminum fixture using a universal testing machine. Single stacks of two to eight snap domes each exhibited very predictable and repeatable buckling behavior (i.e., stack buckling load is simply the sum of individual snap dome buckling loads) when deflected at speeds typical for patients walking with a regular gait. The double stack configuration only works when supporting legs of the opposing snap dome stacks are offset by half the angle between adjacent legs. The lower load stack buckles first, while the higher load stack buckles at its threshold load because of the very low force required to keep the lower load stack collapsed. While the presence of a spacer has little effect on the double stack buckling behavior under controlled rate deflection in a precision test fixture, it was required for proper functioning of a LWBI prototype probably because of looser dimensional tolerances. The type of substrate that snap dome stacks are in contact with has little effect on the buckling loads as long as the material is not too soft. Finally, the speed of deflection within the expected range of ambulating patients has an insignificant effect on the LWBI’s buckling behavior. A LWBI prototype was designed based on the observed characteristics of the snap dome double stack with a spacer plate between the upper and lower load stacks. The prototype was installed in a recess in the insole of a biomechanical shoe beneath the patient’s heel. The shoe with LWBI was tested by various subjects pushing on a force plate and the upper and lower buckling loads were clearly indicated to the subject by audible and tactile click and measured as ground reaction force versus time. Future work will focus on further testing of the device and refinement of the design for various medical appliances.
    • Download: (713.0Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      An Inexpensive Weight Bearing Indicator With Load Range Capability for Rehabilitation of Patients With Lower Extremity Injuries

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/141530
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorDaniel F. Walczyk
    contributor authorWilliam T. Ziomek
    date accessioned2017-05-09T00:34:39Z
    date available2017-05-09T00:34:39Z
    date copyrightSeptember, 2009
    date issued2009
    identifier issn1932-6181
    identifier otherJMDOA4-28006#031001_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/141530
    description abstractThis paper investigates the mechanical behavior and design of a patent pending device called a load range weight bearing indicator (LWBI), which provides upper and lower range indication to patients with lower extremity injuries as part of a partial weight bearing rehabilitation. The LWBI consists of two opposing stacks (a.k.a. double stack) of snap domes—bistable mechanical elements that snap through only when a threshold weight is applied—sandwiched between a load transfer plate and base plate. The mechanical behavior of a LWBI has been characterized by testing single and double stacks of snap domes in a rigid aluminum fixture using a universal testing machine. Single stacks of two to eight snap domes each exhibited very predictable and repeatable buckling behavior (i.e., stack buckling load is simply the sum of individual snap dome buckling loads) when deflected at speeds typical for patients walking with a regular gait. The double stack configuration only works when supporting legs of the opposing snap dome stacks are offset by half the angle between adjacent legs. The lower load stack buckles first, while the higher load stack buckles at its threshold load because of the very low force required to keep the lower load stack collapsed. While the presence of a spacer has little effect on the double stack buckling behavior under controlled rate deflection in a precision test fixture, it was required for proper functioning of a LWBI prototype probably because of looser dimensional tolerances. The type of substrate that snap dome stacks are in contact with has little effect on the buckling loads as long as the material is not too soft. Finally, the speed of deflection within the expected range of ambulating patients has an insignificant effect on the LWBI’s buckling behavior. A LWBI prototype was designed based on the observed characteristics of the snap dome double stack with a spacer plate between the upper and lower load stacks. The prototype was installed in a recess in the insole of a biomechanical shoe beneath the patient’s heel. The shoe with LWBI was tested by various subjects pushing on a force plate and the upper and lower buckling loads were clearly indicated to the subject by audible and tactile click and measured as ground reaction force versus time. Future work will focus on further testing of the device and refinement of the design for various medical appliances.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleAn Inexpensive Weight Bearing Indicator With Load Range Capability for Rehabilitation of Patients With Lower Extremity Injuries
    typeJournal Paper
    journal volume3
    journal issue3
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.3212557
    journal fristpage31001
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2009:;volume( 003 ):;issue: 003
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian