    Studies on Design Optimization of Coronary Stents

    Source: Journal of Medical Devices:;2008:;volume( 002 ):;issue: 001::page 11004
    Author:
    K. Srinivas
    ,
    T. Yamaguchi
    ,
    T. Nakayama
    ,
    M. Ohta
    ,
    S. Obayashi
    DOI: 10.1115/1.2885145
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: The stent design itself seems to be one of the factors responsible for restenosis. As a remedy, the present work attempts to perform a design optimization of coronary stents from a hemodynamic point of view. For the purpose, we have applied the principles of modern exploration of design space restricting ourselves to two-dimensional considerations. Width, thickness, and spacing of the struts of the stent formed the design variables. The objectives chosen for optimization were the vorticity generated, length of recirculation zone, and the reattachment distance in between the struts. Both semicircular and rectangular cross sections of stents were included. Starting with the range of design variables, sample stent cases were generated using Latin hypercube sampling. Objective functions were calculated for each of these by computing the two-dimensional flow using software FLUENT under the assumption of a steady, Newtonian flow considering a model stent with three struts. This was followed by Kriging to construct a response surface, which gives the relationship between the objectives and the design variables. The procedure gave nondominated fronts, which consist of optimized designs. Stents with minimum vorticity, with minimum recirculation distance, and the ones with maximum reattachment length in between struts were generated. The procedure is capable of producing the optimum set of design variables to achieve the prescribed objectives.
    keyword(s): Design , Optimization , Functions , stents , Flow (Dynamics) AND Vorticity ,
    • Download: (270.1Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Studies on Design Optimization of Coronary Stents

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/139091
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorK. Srinivas
    contributor authorT. Yamaguchi
    contributor authorT. Nakayama
    contributor authorM. Ohta
    contributor authorS. Obayashi
    date accessioned2017-05-09T00:30:03Z
    date available2017-05-09T00:30:03Z
    date copyrightMarch, 2008
    date issued2008
    identifier issn1932-6181
    identifier otherJMDOA4-27988#011004_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/139091
    description abstractThe stent design itself seems to be one of the factors responsible for restenosis. As a remedy, the present work attempts to perform a design optimization of coronary stents from a hemodynamic point of view. For the purpose, we have applied the principles of modern exploration of design space restricting ourselves to two-dimensional considerations. Width, thickness, and spacing of the struts of the stent formed the design variables. The objectives chosen for optimization were the vorticity generated, length of recirculation zone, and the reattachment distance in between the struts. Both semicircular and rectangular cross sections of stents were included. Starting with the range of design variables, sample stent cases were generated using Latin hypercube sampling. Objective functions were calculated for each of these by computing the two-dimensional flow using software FLUENT under the assumption of a steady, Newtonian flow considering a model stent with three struts. This was followed by Kriging to construct a response surface, which gives the relationship between the objectives and the design variables. The procedure gave nondominated fronts, which consist of optimized designs. Stents with minimum vorticity, with minimum recirculation distance, and the ones with maximum reattachment length in between struts were generated. The procedure is capable of producing the optimum set of design variables to achieve the prescribed objectives.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleStudies on Design Optimization of Coronary Stents
    typeJournal Paper
    journal volume2
    journal issue1
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.2885145
    journal fristpage11004
    identifier eissn1932-619X
    keywordsDesign
    keywordsOptimization
    keywordsFunctions
    keywordsstents
    keywordsFlow (Dynamics) AND Vorticity
    treeJournal of Medical Devices:;2008:;volume( 002 ):;issue: 001
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian