    Archive

    Radiographic Marking of Patient’s Gravity Line on X-Ray Images

    Source: Journal of Medical Devices:;2008:;volume( 002 ):;issue: 002::page 27546
    Author:
    Chunhui Wu
    ,
    Amir A. Mehbod
    ,
    Ensor Transfeldt
    DOI: 10.1115/1.2932446
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Objective: To develop a clinically feasible apparatus for measuring and marking patient’s gravity line position on spinal x-ray images. Background: Spinal balance of patients with deformity or instability is an important clinical parameter for patient diagnosis, surgery planning and treatment outcome evaluation. Traditional assessment of spinal balance is performed on full spine x-ray images using C7 plumb line relative to central sacral line in the coronal plane or posterior superior corner of S1 in the sagittal plane. Researchers have developed measurement techniques using a force plate synchronized with x-ray equipment and offline x-ray image processing. The complexity of these techniques limits their clinical applications. In this study, we present a new system to mark gravity line directly on an x-ray image. Methods: The system consists of a force plate, a signal conditioner and processor, a motion controller and a motor assembly that moves a radio-opaque marker. The processor calculates the location of gravity line and marker position, and transmits motion commands to the motion controller, which drives a motor assembly and moves the marker to appropriate positions. The acquired x-ray image has a mark indicating the instantaneous gravity line position. Evaluation: A prototype was installed in the radiology department. More than 200 patients were instructed to stand on the force plate. Exposure was performed as in routine x-ray imaging. The system does not affect routine x-ray workflow and provides valuable information for spine surgeons.
    • Download: (27.13Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Radiographic Marking of Patient’s Gravity Line on X-Ray Images

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/139054
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorChunhui Wu
    contributor authorAmir A. Mehbod
    contributor authorEnsor Transfeldt
    date accessioned2017-05-09T00:29:59Z
    date available2017-05-09T00:29:59Z
    date copyrightJune, 2008
    date issued2008
    identifier issn1932-6181
    identifier otherJMDOA4-27991#027546_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/139054
    description abstractObjective: To develop a clinically feasible apparatus for measuring and marking patient’s gravity line position on spinal x-ray images. Background: Spinal balance of patients with deformity or instability is an important clinical parameter for patient diagnosis, surgery planning and treatment outcome evaluation. Traditional assessment of spinal balance is performed on full spine x-ray images using C7 plumb line relative to central sacral line in the coronal plane or posterior superior corner of S1 in the sagittal plane. Researchers have developed measurement techniques using a force plate synchronized with x-ray equipment and offline x-ray image processing. The complexity of these techniques limits their clinical applications. In this study, we present a new system to mark gravity line directly on an x-ray image. Methods: The system consists of a force plate, a signal conditioner and processor, a motion controller and a motor assembly that moves a radio-opaque marker. The processor calculates the location of gravity line and marker position, and transmits motion commands to the motion controller, which drives a motor assembly and moves the marker to appropriate positions. The acquired x-ray image has a mark indicating the instantaneous gravity line position. Evaluation: A prototype was installed in the radiology department. More than 200 patients were instructed to stand on the force plate. Exposure was performed as in routine x-ray imaging. The system does not affect routine x-ray workflow and provides valuable information for spine surgeons.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleRadiographic Marking of Patient’s Gravity Line on X-Ray Images
    typeJournal Paper
    journal volume2
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.2932446
    journal fristpage27546
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2008:;volume( 002 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace