    Design of an Ex Vivo Culture System to Investigate the Effects of Shear Stress on Cardiovascular Tissue

    Source: Journal of Biomechanical Engineering:;2008:;volume( 130 ):;issue: 003::page 35001
    Author:
    Adnan Elhammali
    ,
    Kartik Balachandran
    ,
    Hanjoong Jo
    ,
    Ajit P. Yoganathan
    ,
    Philippe Sucosky
    ,
    Muralidhar Padala
    DOI: 10.1115/1.2907753
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Mechanical forces are known to affect the biomechanical properties of native and engineered cardiovascular tissue. In particular, shear stress that results from the relative motion of heart valve leaflets with respect to the blood flow is one important component of their mechanical environment in vivo. Although different types of bioreactors have been designed to subject cells to shear stress, devices to expose biological tissue are few. In an effort to address this issue, the aim of this study was to design an ex vivo tissue culture system to characterize the biological response of heart valve leaflets subjected to a well-defined steady or time-varying shear stress environment. The novel apparatus was designed based on a cone-and-plate viscometer. The device characteristics were defined to limit the secondary flow effects inherent to this particular geometry. The determination of the operating conditions producing the desired shear stress profile was streamlined using a computational fluid dynamic (CFD) model validated with laser Doppler velocimetry. The novel ex vivo tissue culture system was validated in terms of its capability to reproduce a desired cone rotation and to maintain sterile conditions. The CFD results demonstrated that a cone angle of 0.5deg, a cone radius of 40mm, and a gap of 0.2mm between the cone apex and the plate could limit radial secondary flow effects. The novel cone-and-plate permits to expose nine tissue specimens to an identical shear stress waveform. The whole setup is capable of accommodating four cone-and-plate systems, thus concomitantly subjecting 36 tissue samples to desired shear stress condition. The innovative design enables the tissue specimens to be flush mounted in the plate in order to limit flow perturbations caused by the tissue thickness. The device is capable of producing shear stress rates of up to 650dyncm−2s−1 (i.e., maximum shear stress rate experienced by the ventricular surface of an aortic valve leaflet) and was shown to maintain tissue under sterile conditions for 120h. The novel ex vivo tissue culture system constitutes a valuable tool toward elucidating heart valve mechanobiology. Ultimately, this knowledge will permit the production of functional tissue engineered heart valves, and a better understanding of heart valve biology and disease progression.
    keyword(s): Stress , Shear (Mechanics) , Biological tissues , Computational fluid dynamics , Design , Flow (Dynamics) , Laser Doppler anemometry , Geometry , Cardiovascular system , Valves , Light trucks AND Measurement ,
    • Download: (525.3Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Design of an Ex Vivo Culture System to Investigate the Effects of Shear Stress on Cardiovascular Tissue

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/137450
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorAdnan Elhammali
    contributor authorKartik Balachandran
    contributor authorHanjoong Jo
    contributor authorAjit P. Yoganathan
    contributor authorPhilippe Sucosky
    contributor authorMuralidhar Padala
    date accessioned2017-05-09T00:26:59Z
    date available2017-05-09T00:26:59Z
    date copyrightJune, 2008
    date issued2008
    identifier issn0148-0731
    identifier otherJBENDY-26808#035001_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/137450
    description abstractMechanical forces are known to affect the biomechanical properties of native and engineered cardiovascular tissue. In particular, shear stress that results from the relative motion of heart valve leaflets with respect to the blood flow is one important component of their mechanical environment in vivo. Although different types of bioreactors have been designed to subject cells to shear stress, devices to expose biological tissue are few. In an effort to address this issue, the aim of this study was to design an ex vivo tissue culture system to characterize the biological response of heart valve leaflets subjected to a well-defined steady or time-varying shear stress environment. The novel apparatus was designed based on a cone-and-plate viscometer. The device characteristics were defined to limit the secondary flow effects inherent to this particular geometry. The determination of the operating conditions producing the desired shear stress profile was streamlined using a computational fluid dynamic (CFD) model validated with laser Doppler velocimetry. The novel ex vivo tissue culture system was validated in terms of its capability to reproduce a desired cone rotation and to maintain sterile conditions. The CFD results demonstrated that a cone angle of 0.5deg, a cone radius of 40mm, and a gap of 0.2mm between the cone apex and the plate could limit radial secondary flow effects. The novel cone-and-plate permits to expose nine tissue specimens to an identical shear stress waveform. The whole setup is capable of accommodating four cone-and-plate systems, thus concomitantly subjecting 36 tissue samples to desired shear stress condition. The innovative design enables the tissue specimens to be flush mounted in the plate in order to limit flow perturbations caused by the tissue thickness. The device is capable of producing shear stress rates of up to 650dyncm−2s−1 (i.e., maximum shear stress rate experienced by the ventricular surface of an aortic valve leaflet) and was shown to maintain tissue under sterile conditions for 120h. The novel ex vivo tissue culture system constitutes a valuable tool toward elucidating heart valve mechanobiology. Ultimately, this knowledge will permit the production of functional tissue engineered heart valves, and a better understanding of heart valve biology and disease progression.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleDesign of an Ex Vivo Culture System to Investigate the Effects of Shear Stress on Cardiovascular Tissue
    typeJournal Paper
    journal volume130
    journal issue3
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.2907753
    journal fristpage35001
    identifier eissn1528-8951
    keywordsStress
    keywordsShear (Mechanics)
    keywordsBiological tissues
    keywordsComputational fluid dynamics
    keywordsDesign
    keywordsFlow (Dynamics)
    keywordsLaser Doppler anemometry
    keywordsGeometry
    keywordsCardiovascular system
    keywordsValves
    keywordsLight trucks AND Measurement
    treeJournal of Biomechanical Engineering:;2008:;volume( 130 ):;issue: 003
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian