    Mechanical Properties of a Biodegradable Balloon-expandable Stent From Poly(L-lactide) for Peripheral Vascular Applications

    Source: Journal of Medical Devices:;2007:;volume( 001 ):;issue: 001::page 84
    Author:
    Niels Grabow
    ,
    Carsten M. Bünger
    ,
    Katrin Sternberg
    ,
    Steffen Mews
    ,
    Kathleen Schmohl
    ,
    Klaus-Peter Schmitz
    DOI: 10.1115/1.2355683
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Background: Biodegradable polymeric stents represent a competitive approach to permanent and absorbable metallic stents for vascular applications. Despite major challenges resulting from the mechanical properties of polymeric biomaterials, these stent concepts gain their attraction from their intrinsic potential for controlled biodegradation and facile drug incorporation. This study demonstrates the mechanical properties of a novel balloon-expandable slotted tube stent from PLLA. Method of Approach: Polymeric balloon-expandable slotted tube stents (nominal dimensions: 6.0×25mm) were manufactured by laser machining of solution cast tubes (I.D.=2.8mm, d=270±20μm) from biodegradable (1) PLLA and (2) PLLA/PCL/TEC. The stents were tested in vitro for their mechanical properties: deployment, recoil, shortening, collapse, and creep behavior under a static load of 100mmHg. In vitro degradation was performed in Sørensen buffer solution at 37°C. After 0∕2∕4∕8∕12∕24 weeks the remaining collapse stability and molecular weight were assessed. Results: All stents could be deployed by balloon inflation to 8bar at 1bar∕min (PLLA) and 3bar∕min (PLLA/PCL/TEC). Recoil, shortening, and collapse pressure were: 2.4%∕3.4%∕0.67bar (PLLA), and 8.8%∕2.3%∕0.23bar (PLLA/PCL/TEC). A static load of 100mmHg induced pronounced creep processes in the PLLA/PCL/TEC stent. The PLLA stent remained patent and exhibited no creep propensity. During in vitro degradation an increase in collapse pressure was observed (maxima at 12w: 1.3bar (PLLA), 0.7bar (PLLA/PCL/TEC)). At 24 weeks, molecular weight was decreased by 28% (PLLA), and 52% (PLLA/PCL/TEC). Conclusions: Stents fabricated from pure PLLA exhibited adequate mechanical properties. The slow permissible deployment rate, however, limits their potential application range and demands further development.
    keyword(s): stents AND Collapse ,
    • Download: (200.1Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Mechanical Properties of a Biodegradable Balloon-expandable Stent From Poly(L-lactide) for Peripheral Vascular Applications

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/136602
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorNiels Grabow
    contributor authorCarsten M. Bünger
    contributor authorKatrin Sternberg
    contributor authorSteffen Mews
    contributor authorKathleen Schmohl
    contributor authorKlaus-Peter Schmitz
    date accessioned2017-05-09T00:25:20Z
    date available2017-05-09T00:25:20Z
    date copyrightMarch, 2007
    date issued2007
    identifier issn1932-6181
    identifier otherJMDOA4-27980#84_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/136602
    description abstractBackground: Biodegradable polymeric stents represent a competitive approach to permanent and absorbable metallic stents for vascular applications. Despite major challenges resulting from the mechanical properties of polymeric biomaterials, these stent concepts gain their attraction from their intrinsic potential for controlled biodegradation and facile drug incorporation. This study demonstrates the mechanical properties of a novel balloon-expandable slotted tube stent from PLLA. Method of Approach: Polymeric balloon-expandable slotted tube stents (nominal dimensions: 6.0×25mm) were manufactured by laser machining of solution cast tubes (I.D.=2.8mm, d=270±20μm) from biodegradable (1) PLLA and (2) PLLA/PCL/TEC. The stents were tested in vitro for their mechanical properties: deployment, recoil, shortening, collapse, and creep behavior under a static load of 100mmHg. In vitro degradation was performed in Sørensen buffer solution at 37°C. After 0∕2∕4∕8∕12∕24 weeks the remaining collapse stability and molecular weight were assessed. Results: All stents could be deployed by balloon inflation to 8bar at 1bar∕min (PLLA) and 3bar∕min (PLLA/PCL/TEC). Recoil, shortening, and collapse pressure were: 2.4%∕3.4%∕0.67bar (PLLA), and 8.8%∕2.3%∕0.23bar (PLLA/PCL/TEC). A static load of 100mmHg induced pronounced creep processes in the PLLA/PCL/TEC stent. The PLLA stent remained patent and exhibited no creep propensity. During in vitro degradation an increase in collapse pressure was observed (maxima at 12w: 1.3bar (PLLA), 0.7bar (PLLA/PCL/TEC)). At 24 weeks, molecular weight was decreased by 28% (PLLA), and 52% (PLLA/PCL/TEC). Conclusions: Stents fabricated from pure PLLA exhibited adequate mechanical properties. The slow permissible deployment rate, however, limits their potential application range and demands further development.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleMechanical Properties of a Biodegradable Balloon-expandable Stent From Poly(L-lactide) for Peripheral Vascular Applications
    typeJournal Paper
    journal volume1
    journal issue1
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.2355683
    journal fristpage84
    journal lastpage88
    identifier eissn1932-619X
    keywordsstents AND Collapse
    treeJournal of Medical Devices:;2007:;volume( 001 ):;issue: 001
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian