contributor author | C. Bachmann | |
contributor author | M. Wilson | |
contributor author | V. Kini | |
contributor author | S. Deutsch | |
contributor author | J. M. Tarbell | |
contributor author | A. A. Fontaine | |
date accessioned | 2017-05-09T00:01:50Z | |
date available | 2017-05-09T00:01:50Z | |
date copyright | August, 2000 | |
date issued | 2000 | |
identifier issn | 0148-0731 | |
identifier other | JBENDY-25902#453_1.pdf | |
identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/123342 | |
description abstract | Several types of mechanical cardiac prostheses have been constructed with Delrin occluders, a material that is subject to osmotic swelling. The leaflets are designed to expand to specific tolerances when immersed in blood. The synthetic blood analogs commonly used in vitro contain hydrophilic compounds that can alter the osmotic expansion of the Delrin occluders. A static leak test chamber was employed to illustrate the effects of various test fluids on the sustained regurgitation phase of Delrin valves. [S0148-0731(00)02104-X] | |
publisher | The American Society of Mechanical Engineers (ASME) | |
title | The Osmotic Swelling Characteristics of Cardiac Valve Prostheses | |
type | Journal Paper | |
journal volume | 122 | |
journal issue | 4 | |
journal title | Journal of Biomechanical Engineering | |
identifier doi | 10.1115/1.1286565 | |
journal fristpage | 453 | |
journal lastpage | 454 | |
identifier eissn | 1528-8951 | |
keywords | Fluids | |
keywords | Blood | |
keywords | Prostheses | |
keywords | Valves | |
keywords | Leakage AND Heart valve prostheses | |
tree | Journal of Biomechanical Engineering:;2000:;volume( 122 ):;issue: 004 | |
contenttype | Fulltext | |