A Computer-Controlled Pressure StandardSource: Journal of Fluids Engineering:;1999:;volume( 121 ):;issue: 001::page 210Author:Othon K. Rediniotis
DOI: 10.1115/1.2822006Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: The operating principles, as well as the technical aspects of the implementation of a new computer-controlled pressure standard are presented. The instrument can have dual use: either as a pressure source or as a pressure sensor. The device is intended mostly for use in problems where small differential pressures are of interest, i. e., 0–2.5 KPa and high accuracy is desired. Such a pressure range encompasses, for example, most of the pressure measurement applications in subsonic wind-tunnel testing. The device interfaces to a PC and is ideal for fully-automated pressure transducer calibration applications. The accuracy of the pressures produced or measured by the device is 0.08 percent F. S. (Full Scale).
keyword(s): Pressure , Computers , Calibration , Wind tunnels , Pressure measurement , Pressure transducers , Pressure sensors , Instrumentation AND Testing ,
|
Collections
Show full item record
| contributor author | Othon K. Rediniotis | |
| date accessioned | 2017-05-09T00:00:09Z | |
| date available | 2017-05-09T00:00:09Z | |
| date copyright | March, 1999 | |
| date issued | 1999 | |
| identifier issn | 0098-2202 | |
| identifier other | JFEGA4-27137#210_1.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/122418 | |
| description abstract | The operating principles, as well as the technical aspects of the implementation of a new computer-controlled pressure standard are presented. The instrument can have dual use: either as a pressure source or as a pressure sensor. The device is intended mostly for use in problems where small differential pressures are of interest, i. e., 0–2.5 KPa and high accuracy is desired. Such a pressure range encompasses, for example, most of the pressure measurement applications in subsonic wind-tunnel testing. The device interfaces to a PC and is ideal for fully-automated pressure transducer calibration applications. The accuracy of the pressures produced or measured by the device is 0.08 percent F. S. (Full Scale). | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | A Computer-Controlled Pressure Standard | |
| type | Journal Paper | |
| journal volume | 121 | |
| journal issue | 1 | |
| journal title | Journal of Fluids Engineering | |
| identifier doi | 10.1115/1.2822006 | |
| journal fristpage | 210 | |
| journal lastpage | 212 | |
| identifier eissn | 1528-901X | |
| keywords | Pressure | |
| keywords | Computers | |
| keywords | Calibration | |
| keywords | Wind tunnels | |
| keywords | Pressure measurement | |
| keywords | Pressure transducers | |
| keywords | Pressure sensors | |
| keywords | Instrumentation AND Testing | |
| tree | Journal of Fluids Engineering:;1999:;volume( 121 ):;issue: 001 | |
| contenttype | Fulltext |