    Multiple Minima in Inverse Hyperthermia Temperature Estimation Problems

    Source: Journal of Biomechanical Engineering:;1993:;volume( 115 ):;issue: 003::page 239
    Author:
    C.-T. Liauh
    ,
    R. B. Roemer
    DOI: 10.1115/1.2895481
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Using one-, two-, and three-dimensional numerical simulation models it is shown that multiple minima solutions exist for some inverse hyperthermia temperature estimation problems. This is a new observation that has important implications for all potential applications of these inverse techniques. The general conditions under which these multiple minima occur are shown to be solely due to the existence of symmetries in the bio-heat transfer model used to solve the inverse problem. General rules for determining the number of these global minimum points in the unknown parameter (perfusion) space are obtained for several geometrically symmetric (with respect to the sensor placement and the inverse case blood perfusion model) one-, two- and three-dimensional problem formulations with multiple perfusion regions when no model mismatch is present. As the amount of this symmetry is successively reduced, all but one of these global minima caused by symmetry become local minima. A general approach for (a) detecting when the inverse algorithm has converged to a local minimum, and (b) for using that knowledge to direct the search algorithm toward the global minimum is presented. A three-dimensional, random perfusion distribution example is given which illustrates the effects of the multiple minima on the performance of a state and parameter estimation algorithm. This algorithm attempts to reconstruct the entire temperature field during simulated hyperthermia treatments based on knowledge of measured temperatures from a limited number of locations.
    keyword(s): Temperature , Algorithms , Bioheat transfer , Blood , Inverse problems , Parameter estimation , Sensor placement AND Computer simulation ,
    • Download: (934.4Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Multiple Minima in Inverse Hyperthermia Temperature Estimation Problems

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/111556
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorC.-T. Liauh
    contributor authorR. B. Roemer
    date accessioned2017-05-08T23:40:42Z
    date available2017-05-08T23:40:42Z
    date copyrightAugust, 1993
    date issued1993
    identifier issn0148-0731
    identifier otherJBENDY-25919#239_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/111556
    description abstractUsing one-, two-, and three-dimensional numerical simulation models it is shown that multiple minima solutions exist for some inverse hyperthermia temperature estimation problems. This is a new observation that has important implications for all potential applications of these inverse techniques. The general conditions under which these multiple minima occur are shown to be solely due to the existence of symmetries in the bio-heat transfer model used to solve the inverse problem. General rules for determining the number of these global minimum points in the unknown parameter (perfusion) space are obtained for several geometrically symmetric (with respect to the sensor placement and the inverse case blood perfusion model) one-, two- and three-dimensional problem formulations with multiple perfusion regions when no model mismatch is present. As the amount of this symmetry is successively reduced, all but one of these global minima caused by symmetry become local minima. A general approach for (a) detecting when the inverse algorithm has converged to a local minimum, and (b) for using that knowledge to direct the search algorithm toward the global minimum is presented. A three-dimensional, random perfusion distribution example is given which illustrates the effects of the multiple minima on the performance of a state and parameter estimation algorithm. This algorithm attempts to reconstruct the entire temperature field during simulated hyperthermia treatments based on knowledge of measured temperatures from a limited number of locations.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleMultiple Minima in Inverse Hyperthermia Temperature Estimation Problems
    typeJournal Paper
    journal volume115
    journal issue3
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.2895481
    journal fristpage239
    journal lastpage246
    identifier eissn1528-8951
    keywordsTemperature
    keywordsAlgorithms
    keywordsBioheat transfer
    keywordsBlood
    keywordsInverse problems
    keywordsParameter estimation
    keywordsSensor placement AND Computer simulation
    treeJournal of Biomechanical Engineering:;1993:;volume( 115 ):;issue: 003
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian