| contributor author | S. Evanson | |
| contributor author | M. Otaka | |
| contributor author | K. Hasegawa | |
| date accessioned | 2017-05-08T23:38:37Z | |
| date available | 2017-05-08T23:38:37Z | |
| date copyright | January, 1992 | |
| date issued | 1992 | |
| identifier issn | 0094-4289 | |
| identifier other | JEMTA8-26946#41_1.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/110353 | |
| description abstract | We report on apparatus for the magnetic material characterization of a wide range of steel grades of high and low magnetic permeability. The instrument has been used to detect thermal aging in Duplex steel by the measurement of the residual magnetization (Br ) of test specimens. A method to estimate the value of Br by remote measurements up to 200 mm from a specimen is described. The results are compared with data from conventional hysteresis loop measurements. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | SQUID NDE Apparatus and Magnetic Calibration to Detect Degradation in Duplex Stainless Steel | |
| type | Journal Paper | |
| journal volume | 114 | |
| journal issue | 1 | |
| journal title | Journal of Engineering Materials and Technology | |
| identifier doi | 10.1115/1.2904138 | |
| journal fristpage | 41 | |
| journal lastpage | 45 | |
| identifier eissn | 1528-8889 | |
| keywords | Permeability | |
| keywords | Steel | |
| keywords | Measurement | |
| keywords | Magnetic materials | |
| keywords | Superconducting quantum interference devices | |
| keywords | Nondestructive evaluation | |
| keywords | Magnetization | |
| keywords | Instrumentation | |
| keywords | Calibration AND Stainless steel | |
| tree | Journal of Engineering Materials and Technology:;1992:;volume( 114 ):;issue: 001 | |
| contenttype | Fulltext | |