    A Device for Measuring Very Low Pressure Differences

    Source: Journal of Fluids Engineering:;1980:;volume( 102 ):;issue: 004::page 499
    Author:
    Takeo Nakagawa
    DOI: 10.1115/1.3240733
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: A new technique is proposed for measuring very low pressure differences between two points in a fluid. The device combines an optical velocity measurement with the Hagen-Poiseuille relation to infer the pressure drop. A prototype model has been constructed which measures pressure differences in the range 3.07×10−5 to 1.59×10−3 cm H2 O. The estimated uncertainty of this technique is approximately 4 percent. However, no accurate, independent calibration of this device has yet been achieved.
    keyword(s): Pressure , Fluids , Engineering prototypes , Calibration , Pressure drop , Velocity measurement AND Uncertainty ,
    • Download: (318.1Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      A Device for Measuring Very Low Pressure Differences

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/93443
    Collections
    • Journal of Fluids Engineering

    Show full item record

    contributor authorTakeo Nakagawa
    date accessioned2017-05-08T23:09:00Z
    date available2017-05-08T23:09:00Z
    date copyrightDecember, 1980
    date issued1980
    identifier issn0098-2202
    identifier otherJFEGA4-26964#499_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/93443
    description abstractA new technique is proposed for measuring very low pressure differences between two points in a fluid. The device combines an optical velocity measurement with the Hagen-Poiseuille relation to infer the pressure drop. A prototype model has been constructed which measures pressure differences in the range 3.07×10−5 to 1.59×10−3 cm H2 O. The estimated uncertainty of this technique is approximately 4 percent. However, no accurate, independent calibration of this device has yet been achieved.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleA Device for Measuring Very Low Pressure Differences
    typeJournal Paper
    journal volume102
    journal issue4
    journal titleJournal of Fluids Engineering
    identifier doi10.1115/1.3240733
    journal fristpage499
    journal lastpage501
    identifier eissn1528-901X
    keywordsPressure
    keywordsFluids
    keywordsEngineering prototypes
    keywordsCalibration
    keywordsPressure drop
    keywordsVelocity measurement AND Uncertainty
    treeJournal of Fluids Engineering:;1980:;volume( 102 ):;issue: 004
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian