| description abstract | A new technique is proposed for measuring very low pressure differences between two points in a fluid. The device combines an optical velocity measurement with the Hagen-Poiseuille relation to infer the pressure drop. A prototype model has been constructed which measures pressure differences in the range 3.07×10−5 to 1.59×10−3 cm H2 O. The estimated uncertainty of this technique is approximately 4 percent. However, no accurate, independent calibration of this device has yet been achieved. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | A Device for Measuring Very Low Pressure Differences | |
| type | Journal Paper | |
| journal volume | 102 | |
| journal issue | 4 | |
| journal title | Journal of Fluids Engineering | |
| identifier doi | 10.1115/1.3240733 | |
| journal fristpage | 499 | |
| journal lastpage | 501 | |
| identifier eissn | 1528-901X | |
| keywords | Pressure | |
| keywords | Fluids | |
| keywords | Engineering prototypes | |
| keywords | Calibration | |
| keywords | Pressure drop | |
| keywords | Velocity measurement AND Uncertainty | |
| tree | Journal of Fluids Engineering:;1980:;volume( 102 ):;issue: 004 | |
| contenttype | Fulltext | |