Show simple item record

contributor authorG. A. Parker
contributor authorA. E. Hay
date accessioned2017-05-08T23:02:31Z
date available2017-05-08T23:02:31Z
date copyrightDecember, 1977
date issued1977
identifier issn0022-0434
identifier otherJDSMAA-26048#293_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/89683
description abstractTests are described on a bidirectional fluidic flowmeter for respiratory measurements using the principle first described by Tanney. Good dynamic response and acceptable linearity is achieved over most of the very wide flow range conditions encountered in respiratory flow measurements. A flow range of 380 litres/minute with a signal-to-noise ratio of 150 and a transient rate flow change of 14,500 litres/minute/second were measured for the best flowmeter configuration.
publisherThe American Society of Mechanical Engineers (ASME)
titleA Planar Fluidic Flowmeter Applicable to Respiratory Flows
typeJournal Paper
journal volume99
journal issue4
journal titleJournal of Dynamic Systems, Measurement, and Control
identifier doi10.1115/1.3427122
journal fristpage293
journal lastpage299
identifier eissn1528-9028
treeJournal of Dynamic Systems, Measurement, and Control:;1977:;volume( 099 ):;issue: 004