    Validation of Cardiac Output as Reported by a Permanently Implanted Wireless Sensor

    Source: Journal of Medical Devices:;2016:;volume( 010 ):;issue: 001::page 11001
    Author:
    Tree, Michael
    ,
    White, Jason
    ,
    Midha, Prem
    ,
    Kiblinger, Samantha
    ,
    Yoganathan, Ajit
    DOI: 10.1115/1.4031799
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: The CardioMEMS heart failure (HF) system was tested for cardiac output (CO) measurement accuracy using an in vitro mock circulatory system. A software algorithm calculates CO based on analysis of the pressure waveform as measured from the pulmonary artery, where the CardioMEMS system resides. Calculated CO was compared to that from reference flow probe in the circulatory system model. CO measurements were compared over a clinically relevant range of stroke volumes and heart rates with normal, pulmonary hypertension (PH), decompensated left heart failure (DLHF), and combined DHLF + PH hemodynamic conditions. The CardioMEMS CO exhibited minimal fixed and proportional bias.
    • Download: (881.1Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Validation of Cardiac Output as Reported by a Permanently Implanted Wireless Sensor

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/161979
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorTree, Michael
    contributor authorWhite, Jason
    contributor authorMidha, Prem
    contributor authorKiblinger, Samantha
    contributor authorYoganathan, Ajit
    date accessioned2017-05-09T01:31:39Z
    date available2017-05-09T01:31:39Z
    date issued2016
    identifier issn1932-6181
    identifier othermed_010_01_011001.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/161979
    description abstractThe CardioMEMS heart failure (HF) system was tested for cardiac output (CO) measurement accuracy using an in vitro mock circulatory system. A software algorithm calculates CO based on analysis of the pressure waveform as measured from the pulmonary artery, where the CardioMEMS system resides. Calculated CO was compared to that from reference flow probe in the circulatory system model. CO measurements were compared over a clinically relevant range of stroke volumes and heart rates with normal, pulmonary hypertension (PH), decompensated left heart failure (DLHF), and combined DHLF + PH hemodynamic conditions. The CardioMEMS CO exhibited minimal fixed and proportional bias.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleValidation of Cardiac Output as Reported by a Permanently Implanted Wireless Sensor
    typeJournal Paper
    journal volume10
    journal issue1
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4031799
    journal fristpage11001
    journal lastpage11001
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2016:;volume( 010 ):;issue: 001
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian