| contributor author | Tree, Michael | |
| contributor author | White, Jason | |
| contributor author | Midha, Prem | |
| contributor author | Kiblinger, Samantha | |
| contributor author | Yoganathan, Ajit | |
| date accessioned | 2017-05-09T01:31:39Z | |
| date available | 2017-05-09T01:31:39Z | |
| date issued | 2016 | |
| identifier issn | 1932-6181 | |
| identifier other | med_010_01_011001.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/161979 | |
| description abstract | The CardioMEMS heart failure (HF) system was tested for cardiac output (CO) measurement accuracy using an in vitro mock circulatory system. A software algorithm calculates CO based on analysis of the pressure waveform as measured from the pulmonary artery, where the CardioMEMS system resides. Calculated CO was compared to that from reference flow probe in the circulatory system model. CO measurements were compared over a clinically relevant range of stroke volumes and heart rates with normal, pulmonary hypertension (PH), decompensated left heart failure (DLHF), and combined DHLF + PH hemodynamic conditions. The CardioMEMS CO exhibited minimal fixed and proportional bias. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | Validation of Cardiac Output as Reported by a Permanently Implanted Wireless Sensor | |
| type | Journal Paper | |
| journal volume | 10 | |
| journal issue | 1 | |
| journal title | Journal of Medical Devices | |
| identifier doi | 10.1115/1.4031799 | |
| journal fristpage | 11001 | |
| journal lastpage | 11001 | |
| identifier eissn | 1932-619X | |
| tree | Journal of Medical Devices:;2016:;volume( 010 ):;issue: 001 | |
| contenttype | Fulltext | |